FDA Adverse Event Malfunction Summary report: N

LEKSELL GAMMAPLAN

MDR report key: 23877055 · Received December 23, 2025

Report

Report Number
3015232217-2025-00063
Event Type
Malfunction
Date Received
December 23, 2025
Report Date
December 17, 2025
Manufacturer
ELEKTA SOLUTIONS AB
Product Code
MUJ
UDI-DI
07340201502136
PMA / PMN Number
K232854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DUE TO AN ADMINISTRATIVE ERROR, THE REPORT WAS ORIGINALLY SUBMITTED UNDER ., WHICH WAS AN INCORRECT/INCOMPLETE REFERENCE NUMBER. THE CORRECT REFERENCE SHOULD HAVE BEEN .. THEREFORE, THIS REPORT HAS BEEN SUBMITTED AS AN INITIAL AND FINAL TO CORRECT THE REFERENCE NUMBER. H6 UPDATED. H11 UPDATED: THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. THE CUSTOMER REPORTED THAT AN OBSOLETE STEREOTACTIC REFERENCE WAS USED FOR TREATMENT. IT WAS ASCERTAINED THAT A TREATMENT PLAN WAS CREATED USING FRAMELESS MRI IMAGES, G-FRAME FIXATION, AND CONE BEAM COMPUTED TOMOGRAPHY (CBCT) AS THE STEREOTACTIC REFERENCE. DURING THE PLANNING PROCESS, THE CUSTOMER ELECTED TO REPOSITION THE FRAME AND ACQUIRED A NEW CBCT. HOWEVER, DUE TO A USE ERROR, THE TREATMENT PLAN WAS NOT CORRECTLY UPDATED TO REFLECT THE NEW FRAME POSITION. AS A RESULT, THE TREATMENT WAS DELIVERED BASED ON THE PREVIOUS FRAME PLACEMENT. THE PLAN HAD BEEN APPROVED AND PARTIALLY DELIVERED BEFORE THE OPERATOR IDENTIFIED THE DISCREPANCY AND PAUSED THE TREATMENT. THE SYSTEM WORKED AS DESIGNED AND INTENDED. ELEKTA HAVE NOT BEEN PROVIDED WITH INFORMATION REGARDING THE PATIENT'S CONDITION. AN IMPORTANT FIELD SAFETY NOTICE (100-01-102-019) WILL BE SENT TO ALL AFFECTED CUSTOMERS FROM 17 DECEMBER 2025 (ELEKTA REFERENCE #: (B)(4)).

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT AN OBSOLETE STEREOTACTIC REFERENCE USED FOR TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1800762 LEKSELL GAMMAPLAN RADIONUCLIDE RT TREATMENT PLANNING SYSTEM MUJ ELEKTA SOLUTIONS AB 11.4.1 07340201502136

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown