FDA Adverse Event Malfunction Summary report: N

ELECSYS ANTI-TSHR

MDR report key: 23876444 · Received December 23, 2025

Report

Report Number
1823260-2025-05304
Event Type
Malfunction
Date Received
December 23, 2025
Date of Event
December 8, 2025
Report Date
February 6, 2026
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JZO
UDI-DI
07613336160329
PMA / PMN Number
K193313
Removal / Correction Number
RES 98103
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ISSUES HAVE BEEN IDENTIFIED WITH THE ELECSYS ANTI-TSHR ASSAY, PARTICULARLY AROUND THE 1.75 IU/L MEDICAL DECISION POINT. ROCHE DIAGNOSTICS HAS ISSUED AN URGENT MEDICAL DEVICE CORRECTION FOR THIS ISSUE, ENTITLED ""ELECSYS® ANTI-TSHR ASSAY ¿ DEVIATIONS ON COBAS® E 411, 402, AND 801 ANALYZERS."" THIS CORRECTION HAS BEEN REPORTED TO FDA. ROCHE DIAGNOSTICS INVESTIGATED AND CONFIRMED THE CUSTOMERS' ALLEGATIONS. THE ROOT CAUSE FOR THE VARIATIONS IN THE ELECSYS ANTI-TSHR ASSAY WAS AN INSUFFICIENCY OF THE ASSAY'S ROUTINE PRODUCTION PROCESS. FOR THE PLATFORM-TO-PLATFORM DEVIATION (COBAS E 411 VS. COBAS E 402/E 801), DIFFERENCES AROSE DUE TO AN INSUFFICIENT ALIGNMENT CONCEPT BETWEEN THE PLATFORMS. THESE VARIATIONS WERE FURTHER INFLUENCED BY UPDATES TO THE ROUTINE PRODUCTION PROCESS FOR THIS ASSAY. FOR THE LOT-TO-LOT DEVIATION ON E 411, MEASURES TAKEN TO ENHANCE ROBUSTNESS LED TO ADDITIONAL SHIFTS. CUSTOMERS WERE ADVISED TO IMMEDIATELY DISCONTINUE USING AND DISCARD THE AFFECTED LOT NUMBERS: 840183 AND 874011 ON THE COBAS E402/E801 ANALYZER. 840177 ON THE COBAS E411 ANALYZER. CUSTOMERS WERE ADVISED TO SWITCH TO THE NEW UNAFFECTED LOT NUMBERS: 908953 ON THE COBAS E402/E801 ANALYZER. 906872 ON THE COBAS E411 ANALYZER. THE COBAS E601 AND E602 MODULES ARE NOT AFFECTED AND CAN BE USED WITH ALL CURRENTLY AVAILABLE LOTS AND WITH THE UPCOMING LOTS WITHOUT RESTRICTIONS.

Additional Manufacturer Narrative · 0

THE COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS ANTI-TSHR RESULTS FOR QC MATERIAL AND PATIENT SAMPLES FROM THE COBAS E 801 ANALYTICAL UNIT. FOR ONE PATIENT, THE INITIAL RESULT WAS 4.7 IU/L. ON (B)(6) 2025, THE REPEAT RESULTS USING A NEW REAGENT CASSETTE OF THE SAME LOT NUMBER WERE 3.0-3.2 IU/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2477120 ELECSYS ANTI-TSHR THYROID AUTOANTIBODY TEST JZO ROCHE DIAGNOSTICS 84018301 07613336160329

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown