FDA Adverse Event Malfunction Summary report: N

TRUMPF MEDICAL

MDR report key: 23872421 · Received December 22, 2025

Report

Report Number
3007143268-2025-00093
Event Type
Malfunction
Date Received
December 22, 2025
Date of Event
November 11, 2025
Report Date
January 12, 2026
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
GDC
UDI-DI
00887761979390
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. ALL ADDITIONAL AND RELEVANT INFORMATION THAT IS IDENTIFIED FOLLOWING COMPLETION OF THE INVESTIGATION WILL BE SUBMITTED IN A FINAL REPORT.

Additional Manufacturer Narrative · 0

THE OPERATING TABLE INVOLVED WAS INSPECTED BY A BAXTER FIELD SERVICE TECHNICIAN. THE ALLEGATION OF HYDRAULIC (FOOT) STAMP WAS DISENGAGING ON ITS OWN COULD BE CONFIRMED. BAXTER EXCHANGED THE OPERATING TABLE AT THE CUSTOMER LOCATION FOR FURTHER INVESTIGATION AT THE MANUFACTURERS PLANT. DURING THIS INVESTIGATION THE CAUSE WAS TRACED TO THE HYDRAULIC UNIT. AFTER THE REPLACEMENT OF THIS PART, THE DEVICE WAS WORKING AS INTENDED. THE ROOT CAUSE OF THE HARDWARE DEFECT WAS NOT IDENTIFIED YET. THE ALLEGED FAILURE MODE WILL BE MONITORED FURTHER VIA POST-MARKET SURVEILLANCE ACTIVITIES AND THE INVESTIGATION FOR THE ROOT CAUSE WILL PROCEED FURTHER.

Description of Event or Problem · 0

ON 25-NOV-2025, A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT PST 500 (PRODUCT CODE 4080100, SERIAL NUMBER (B)(6)), HAD AN ISSUE, THE BRAKES DO NOT EXTEND EVENLY, CAUSING THE OPERATING TABLE TO WOBBLE. ONE BRAKE RELEASES INTRAOPERATIVELY, CAUSING THE OPERATING TABLE TO WOBBLE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR DEATH REPORTED WITH THIS EVENT.

Description of Event or Problem · 0

ON (B)(6) 2025, A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT PST 500 (PRODUCT CODE 4080100, SERIAL NUMBER (B)(6), HAD AN ISSUE, THE BRAKES DO NOT EXTEND EVENLY, CAUSING THE OPERATING TABLE TO WOBBLE. ONE BRAKE RELEASES INTRAOPERATIVELY, CAUSING THE OPERATING TABLE TO WOBBLE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR DEATH REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1564085 TRUMPF MEDICAL TABLE, OPERATING-ROOM, ELECTRICAL GDC BAXTER HEALTHCARE CORPORATION 4080100 N/A 00887761979390

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown