FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SERUM

MDR report key: 23872146 · Received December 22, 2025

Report

Report Number
1024879-2025-02275
Event Type
Malfunction
Date Received
December 22, 2025
Date of Event
November 28, 2025
Report Date
January 16, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903678120
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED TEN PHOTOS FOR INVESTIGATION. THE PHOTOS SHOW MULTIPLE TUBES WITH STOPPERS STUCK INSIDE AND STOPPERS SEPARATED FROM THEIR SHIELDS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOTS, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE LOT 5042703, FOR THE INDICATED FAILURE MODE: CLOSURE SEPARATION. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. BASED ON AN EVALUATION OF SEVERITY AND FREQUENCY IT WAS DETERMINED THAT NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING BD VACUTAINER® SERUM ON AN AUTOMATED ANALYZER, THE STOPPER REMAINED IN THE TUBE WHILE THE SHIELD DETACHED ON ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING BD VACUTAINER® SERUM ON AN AUTOMATED ANALYZER, THE STOPPER REMAINED IN THE TUBE WHILE THE SHIELD DETACHED ON ONE (1) DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1715611 BD VACUTAINER® SERUM TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 5042703 30382903678120

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown