FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR ANALYZER

MDR report key: 2386933 · Received December 22, 2011

Report

Report Number
1628664-2011-00749
Event Type
Malfunction
Date Received
December 22, 2011
Date of Event
December 8, 2011
Report Date
December 14, 2011
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
NIG
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT I2000SR ANALYZER, LIST 03M74-01, (B)(4) MANUFACTURING SITE TO ARCHITECT CA 19-9XR ASSAY, LIST 02K91-25, (B)(4) MANUFACTURING SITE. MDR NUMBERS 1415939-2012-00055 AND 1415939-2012-00056 HAVE BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THOSE MDR NUMBERS.

Additional Manufacturer Narrative · 1

AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE. PATIENT PROBLEM CODE: NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT (B)(4). DEVICE PROBLEM CODE: INCORRECT OR INADEQUATE RESULT (B)(4). COMPLETE PATIENT IDENTIFICATION NUMBER: (B)(4). ARCH I2000SR LN: (B)(4), SN: (B)(4).

Description of Event or Problem · 1

THE CUSTOMER OBSERVED THE FOLLOWING RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ARCHITECT CA 19-9XR ASSAY: INITIAL RESULT OF >1200 U/ML (NEAT; REAGENT LOT (B)(4)) ON ARCHITECT I2000SR SN: (B)(4), RETESTED AT 905.76 U/ML (1:2; REAGENT LOT (B)(4)) ON ARCHITECT I2000SR SN: (B)(4). THE FOLLOWING RESULTS WERE ALL GENERATED WITH REAGENT LOT (B)(4) ON ARCHITECT I2000SR SN: (B)(4): 677.79 U/ML (1:10); >1200 U/ML (NEAT); 702.84 U/ML ; 495.55 U/ML (1:5); 596.74 U/ML (1:10). THE FOLLOWING DAY THE SAMPLE WAS RETESTED AGAIN: >1200 U/ML (NEAT); AND, 603.93 U/ML (1:10). NO SUSPECT RESULTS WERE REPORTED FROM THE LAB. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000SR ANALYZER NIG ABBOTT MANUFACTURING INC

Patients

Seq Age Sex Outcome Treatment
1 ARCHCA 19-9XR LN: (B)(4) LOT: (B)(4)| ARCHCA 19-9XR LN: (B)(4) LOT:01083M500| ARCHCA 19-9XR LN: (B)(4) LOT: (B)(4)| ARCHCA 19-9XR LN: (B)(4) LOT:02949M500| ARCH I2000SR LN: (B)(4) SN: (B)(4)