FDA Adverse Event
Malfunction
Summary report: N
BD MAX¿ SYSTEM, BD MAX¿ INSTRUMENT
MDR report key: 23867554
·
Received December 22, 2025
Report
- Report Number
- 1119779-2025-05525
- Event Type
- Malfunction
- Date Received
- December 22, 2025
- Date of Event
- December 4, 2025
- Report Date
- December 10, 2025
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- OOI
- UDI-DI
- 38290441916098
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
D.2. ADDITIONAL MEDICAL DEVICE TYPES: NQX, NJR, OZN. E.1. INITIAL REPORTER PHONE #: (B)(6). E.1. INITIAL REPORTER FAX #: (B)(6). G.4. THERE WERE MULTIPLE PMA/510K NUMBERS: K111860, K120138, K130470. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem · 0
IT WAS REPORTED THAT DURING USE OF BD MAX¿ SYSTEM, BD MAX¿ INSTRUMENT, A FALSE NEGATIVE MDR-TB PATIENT RESULT WAS OBTAINED. SAMPLE WAS ALSO RUN ON BD BACTEC¿ AND MDR-TB WAS DETECTED. THERE WAS NO REPORT OF PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1605486 | BD MAX¿ SYSTEM, BD MAX¿ INSTRUMENT | THERMAL CYCLER NUCLEIC ACID AMPLIFICATION ANALYSER IVD, LABORATORY | OOI | BECTON, DICKINSON & CO. (SPARKS) | 38290441916098 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |