FDA Adverse Event Injury Summary report: N

MEDLINE

MDR report key: 23865321 · Received December 22, 2025

Report

Report Number
1417592-2025-00733
Event Type
Injury
Date Received
December 22, 2025
Date of Event
November 19, 2025
Report Date
April 8, 2026
Manufacturer
MEDLINE INDUSTRIES, LP
Product Code
BTT
UDI-DI
10197344009589
PMA / PMN Number
K131912
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

UPDATED D9: DEVICE AVAILABLE FOR EVALUATION AND DATE RETURNED TO MANUFACTURER. UPDATED H3: DEVICE EVALUATED BY MANUFACTURER. UPDATED H6: INVESTIGATION FINDINGS (C). UPDATED H6: INVESTIGATION CONCLUSIONS (D).

Additional Manufacturer Narrative · 0

ACCORDING TO THE FACILITY, ON 11/19/2025, WHEN PATIENT WAS USING COMFORT FLOW HUMIDIFICATION SYSTEM, THERE WAS A "SUDDEN BURST OF WATER AND PRESSURE" AND THE "PATIENT STARTED CHOKING AND COUGHING AS WATER SPRAYED IN FACE." THE NURSE "REMOVED THE PATIENT FROM THE HHFNC AND PLACED ON 15 LPM NRB." THE PATIENT'S OXYGEN SATURATION "DROPPED TO 60%, PT RECOVERED AFTER 3 MINUTES." THE FACILITY STATED THE UNIT "TUBING WAS FULL OF WATER." THE PRODUCT HAS BEEN REQUESTED FOR EVALUATION. IT HAS BEEN DETERMINED THAT THE REPORTED EVENT CAUSED OR CONTRIBUTED TO SERIOUS INJURY REQUIRING MEDICAL INTERVENTION, THEREFORE, THIS MEDWATCH IS BEING FILED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED OR OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED. MANUFACTURING DATES FOR REPORTED LOT NUMBERS: 24I0A690- 09/2024, 24L0A197-12/2024, 25A0A623-01/2025. EXPIRATION DATES FOR REPORTED LOT NUMBERS: 24I0A690- 09/2029, 24L0A197-12/2029, 25A0A623-01/2030.

Description of Event or Problem · 0

ACCORDING TO THE FACILITY, ON (B)(6) 2025, WHEN PATIENT WAS USING COMFORT FLOW HUMIDIFICATION SYSTEM, THERE WAS A "SUDDEN BURST OF WATER AND PRESSURE" AND THE "PATIENT STARTED CHOKING AND COUGHING AS WATER SPRAYED IN FACE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183992 MEDLINE HUMIDIFICATION SYSTEM, COMFORT FLO BTT MEDLINE INDUSTRIES, LP 24I0A690, 24L0A197, 25A0A623 10197344009589

Patients

Seq Age Sex Outcome Treatment
1 57 YR Male Other