I-STAT ACT CELITE CARTRIDGE
Report
- Report Number
- 2245578-2011-00359
- Event Type
- Malfunction
- Date Received
- November 17, 2011
- Date of Event
- September 21, 2011
- Report Date
- October 27, 2011
- Manufacturer
- ABBOTT POINT OF CARE, INC.
- Product Code
- JBP
- PMA / PMN Number
- K992571
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). LOT: S11154, MANUFACTURED: 06/2011; EXPIRATION: 11/28/2011.
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT ACT CELITE CARTRIDGES THAT YIELDED A DISCREPANT RESULT ON A PT. CUSTOMER USED TWO DIFFERENT LOTS. (B)(6) 2011: I-STAT ACT 177 AT 0846 (LOT S11196B), I-STAT ACT 181 AT 0901 (LOT S11196B), I-STAT ACT 188 AT 0901 (LOT S11154), I-STAT ACT 195 AT 1022 (LOT S11196B); I-STAT ACT 170 AT 0953 (LOT S11196B). BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. NEW INFORMATION WAS RECEIVED (B)(6) 2011 STATING THAT THERE WAS NO IMPACT TO THE PT AND THERE WAS NO DELAY. HOWEVER, IT IS NOT CLEAR IF PT MANAGEMENT WAS IMPACTED. THE INVESTIGATION WAS COMPLETED ON (B)(6) 2011. FINISH GOODS AND RETAIN TESTING SHOWED THAT LOTS S11196B AND S11154 ARE PERFORMING TO SPECIFICATION. (B)(6) 2011 SUGGESTED THAT PT MANAGEMENT MAY HAVE BEEN IMPACTED DUE TO THE AMOUNT OF HEPARIN AND PROTAMINE ADMINISTERED TO PT. THOUGH THIS IS UNKNOWN BASED ON THE INFORMATION AVAILABLE; 5000 UNITS OF HEPARIN AT 08:30, ACT AT 08:46 - 177; 2000 UNITS OF HEPARIN AT 08:51, ACT AT 09:01 - 1181; 3000 UNITS OF HEPARIN AT 09:06, 50 NG OF PROTAMINE AT 09:30; 7000 UNITS OF HEPARIN AT 09:44, ACT AT 09:53 - 170; 4000 UNITS OF HEPARIN AT 09:57, ACT AT 10:03 TEST CANCELED BY OPERATOR AT 800 SEC. METER DISPLAYED 0. ACT AT 10:22 - 195; 50 NG PROTAMINE AT 10:30 OR 10:40.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT ACT CELITE CARTRIDGE | ACT CELITE CARTRIDGE | JBP | ABBOTT POINT OF CARE, INC. | NA | S11196B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR |