CEREGLIDE
Report
- Report Number
- 3007628272-2025-00063
- Event Type
- Malfunction
- Date Received
- December 19, 2025
- Date of Event
- December 5, 2025
- Report Date
- February 5, 2026
- Manufacturer
- CERENOVUS, INC.
- Product Code
- NRY
- UDI-DI
- 10886704086444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THE INVESTIGATIONAL FINDING OF THE RETURNED DEVICE. THE COMPLAINT PRODUCT WAS RETURNED AND RECEIVED FOR EVALUATION AND ANALYSIS. THE INVESTIGATION IS DOCUMENTED BELOW. INVESTIGATION SUMMARY: A NON-STERILE 144 CM CEREGLIDE 42 CATHETER WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. VISUAL INSPECTION WAS PERFORMED. A KINK WAS NOTED NEXT TO THE ID BAND. NO ADDITIONAL DAMAGES WERE NOTED. THE CATHETER WAS CONNECTED TO A ROTATING HEMOSTATIC VALVE (RHV). THE VALVE WAS TIGHTENED AND NO DAMAGES NOR MALFUNCTIONS WERE IDENTIFIED. THE RHV AND CATHETER WERE FLUSHED TO IDENTIFY MISALIGNMENTS IN THE INTERACTION BETWEEN COMPONENTS; HOWEVER, NO WATER LEAKAGE WAS IDENTIFIED. THE REPORTED ISSUE IN THE COMPLAINT COULD NOT BE CONFIRMED SINCE NO DAMAGES WERE FOUND ON THE DEVICE. OTHER CLINICAL OR PROCEDURAL FACTORS, WHICH COULD NOT BE REPLICATED DURING THE ANALYSIS, MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. THERE IS NO INDICATION THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE MANUFACTURING OF THE DEVICE. THE KINK AT THE ID BAND WAS NOT ORIGINALLY REPORTED, AND THE EXACT TIME OF OCCURRENCE CANNOT BE DETERMINED; THEREFORE, THIS IS NOT CONSIDERED RELATED TO THE ISSUE REPORTED. A REVIEW OF MANUFACTURING DOCUMENTATION ASSOCIATED WITH THIS LOT (31766627) PRESENTED NO ISSUES DURING THE MANUFACTURING OR INSPECTION PROCESSES RELATED TO THE REPORTED COMPLAINT. THERE WERE NO INTERNAL ACTIONS RELATED TO DEVICE MANUFACTURE OR INSPECTION. AS PART OF JOHNSON & JOHNSON MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED AT A LATER TIME. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. UPDATED SECTIONS: B.4, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, GENDER, AND WEIGHT, WERE NOT PROVIDED. SECTION D.2B: PROCODE IS NRY/QJP. SECTION E.1: THE INITIAL REPORTER PHONE IS NOT AVAILABLE / REPORTED. SECTION H.3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. THE PRODUCT ANALYSIS TEAM REVIEWED THE PRE-SHIPMENT PHOTOS OF THE COMPLAINT DEVICE THAT WERE ADDED TO THE COMPLAINT FILE ON 15-DEC-2025. THE REVIEW IS DOCUMENTED BELOW. [PHOTO REVIEW]: IN THE PHOTOS, A ROTATING HEMOSTASIS VALVE (RHV) WITH EXTENSION TUBING WAS CONNECTED TO A CEREGLIDE 42 CATHETER. MAGNIFIED OBSERVATIONS WERE PERFORMED USING A MICROSCOPE, AND NO ANOMALIES WERE FOUND IN APPEARANCE. THE HUB WAS INSPECTED, AND NO ANOMALIES WERE FOUND. ONE KINK WAS FOUND NEAR THE ID BAND. DURING THE VALVE TIGHTENING PROCESS, THE VALVE AS ABLE TO BE TIGHTENED ALL THE WAY, THE CONNECTION BETWEEN THE VALVE AND THE HUB REMAINED INTACT AND NO DETACHMENT OCCURRED. THE KINK NEXT TO THE ID BAND WAS NOT ORIGINALLY REPORTED, AND THE EXACT TIME OF OCCURRENCE CANNOT BE DETERMINED; THEREFORE, THIS IS NOT CONSIDERED RELATED TO THE ISSUE REPORTED. THE REPORTED ISSUE IN THE COMPLAINT CANNOT BE CONFIRMED; HOWEVER, THIS INVESTIGATION WAS PERFORMED BASED ONLY ON THE PHOTO PROVIDED. AN ASSESSMENT WILL BE PERFORMED AS PER THE CONDITIONS OF THE DEVICE RETURNED. A REVIEW OF MANUFACTURING DOCUMENTATION ASSOCIATED WITH THIS LOT (31766627) PRESENTED NO ISSUES DURING THE MANUFACTURING OR INSPECTION PROCESSES RELATED TO THE REPORTED COMPLAINT. THERE WERE NO INTERNAL ACTIONS RELATED TO DEVICE MANUFACTURE OR INSPECTION. AS PART OF J&J MEDTECH QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY CERENOVUS, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, CERENOVUS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT THE PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE ON 26-JAN-2026. A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED ONCE THE PRODUCT INVESTIGATION HAS BEEN COMPLETED. UPDATED SECTIONS: B.4, D.9, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A MECHANICAL THROMBECTOMY TARGETING AN OCCLUSION IN THE M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) TO TREAT AN ISCHEMIC STROKE, THE ISSUES WITH THE 144 CM CEREGLIDE 42 CATHETER (NIC42144C / 31766627) AND THE 132CM CEREGLIDE 57 CATHETER (NIC57132C / 31766630) WERE NOTICED DURING THE PROCEDURE PREPARATION. THE DEVICES WERE USED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFUS). IT WAS REPORTED THAT ¿WHEN THE PRODUCT-SUPPLIED LUER VALVE WAS ATTACHED TO TWO CEREGLIDE CATHETERS DURING THE USUAL PREPARATION, BOTH EXHIBITED A SENSATION OF NOT BEING ABLE TO BE FULLY SCREWED ON; EVEN WHEN IT SEEMED THEY HAD BEEN TIGHTENED COMPLETELY, THEY APPEARED TO LOOSEN SLIGHTLY ONCE RELEASED.¿ THE PROCEDURE WAS CONTINUED WITH SOME CONCERN; IT WAS SUCCESSFULLY COMPLETED WITHOUT ANY NEGATIVE IMPACT ON THE PATIENT. THE PHYSICIAN COULD NOT DETERMINE THE CAUSE OF THE ISSUE AND REQUESTED AN INVESTIGATION. THE DEVICES WERE INSPECTED BY THE CERENOVUS SALES REPRESENTATIVE, THE SAME ISSUE WAS EXPERIENCED ¿ THE INABILITY TO FULLY SCREW THE VALVE ON, WHICH IS AN OCCURRENCE THAT HAS NOT BEEN PREVIOUSLY ENCOUNTERED. ON 10-DEC-2025, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, BOTH CATHETERS WERE USED FOR THE THROMBECTOMY PROCEDURE EVEN THOUGH THERE WAS THE REPORTED ISSUE WITH THE LUER VALVE. THE INFORMATION INDICATED THAT BOTH CEREGLIDE 42 AND CEREGLIDE 57 WERE USED BUT IT WAS NOT SPECIFIED WHICH OF THE TWO WAS USED TO COMPLETE THE PROCEDURE. ON 14-DEC-2025, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE PATIENT IS A JAPANESE FEMALE IN HER 70S. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ON 15-DEC-2025, PRE-SHIPMENT PHOTOS OF THE COMPLAINT DEVICE WERE ADDED TO THE COMPLAINT FILE. THE PHOTOS WILL BE REVIEWED BY THE PRODUCT ANALYSIS TEAM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 398777 | CEREGLIDE | CATHETER, THROMBUS RETRIEVER CATHETER, PERCUTANEOUS, NEUROVASCULATURE | NRY | CERENOVUS, INC. | 31766627 | 10886704086444 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | TREVO TRAK® MICROCATHETER (STRYKER) |