FDA Adverse Event Malfunction Summary report: N

THREADED IM NAIL 1.8MM, 50MM, TI

MDR report key: 23844813 · Received December 18, 2025

Report

Report Number
2031009-2025-00027
Event Type
Malfunction
Date Received
December 18, 2025
Date of Event
November 21, 2025
Report Date
December 18, 2025
Manufacturer
TRIMED, INC.
Product Code
HWC
UDI-DI
00842188124051
PMA / PMN Number
K211783
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 0

SALES MANAGER REPORTED ON (B)(6) 2025 THAT AN IMN1.8-50 T BROKE. THE REP IN THE CASE STATED THAT THE SURGEON REAMED WITH THE APPROPRIATE DRILL PRIOR TO PUTTING IN THE NAIL BEYOND WHERE THE NAIL GOT STUCK. AS HE WAS ADVANCING THE NAIL HE STARTED GETTING RESISTANCE AND TOOK AN X-RAY AND SAW THE NAIL WAS BENDING AND TRIED TO BACK IT OUT AT WHICH POINT THE NAIL BROKE. THE SURGEON SUCCESSFULLY REMOVED ALL PIECES OF THE BROKEN DEVICE AND USED A DIFFERENT NAIL TO COMPLETE THE SURGERY. THE CHANGE IN INVENTORY CAUSED ABOUT A 30-40-MINUTE DELAY IN SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2387118 THREADED IM NAIL 1.8MM, 50MM, TI IMN1.8-50 T HWC TRIMED, INC. IMN1.8-50 T 00842188124051

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other