FDA Adverse Event Malfunction Summary report: N

Q-SYTE

MDR report key: 23839880 · Received December 18, 2025

Report

Report Number
1710034-2025-01968
Event Type
Malfunction
Date Received
December 18, 2025
Date of Event
August 20, 2025
Report Date
December 8, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FPA
UDI-DI
00382903851003
PMA / PMN Number
K013621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385100 AND LOT NUMBER 5031461. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT SIMILAR ISSUES DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

WHILE CONNECTING THE CONNECTOR DURING AN IV INFUSION FOR A PATIENT, THE CONNECTOR LEAKED FLUID. REPLACED WITH A NEW CONNECTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2385813 Q-SYTE SET, ADMINISTRATION, INTRAVASCULAR FPA BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5031461 00382903851003

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown