HAMILTON-C6
Report
- Report Number
- 3001421318-2025-00978
- Event Type
- Malfunction
- Date Received
- December 18, 2025
- Date of Event
- December 2, 2025
- Report Date
- May 20, 2026
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002808590
- PMA / PMN Number
- K201658
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
HAMILTON MEDICAL AGS REFERENCE: (B)(4). HAMILTON MEDICAL AGS CONCLUSION: INVESTIGATION IS ONGOING.
HAMILTON MEDICAL AGS REFERENCE: (B)(4); HAMILTON MEDICAL AGS FOLLOW UP: IT WAS REPORTED THAT THE ACTIVE HUMIDIFICATION WAS FOUND SWITCHED OFF AND THAT THE BREATHING CIRCUIT FILLED WITH WATER AFTER REACTIVATION.THE LOGFILE ANALYSIS SHOWED NO TECHNICAL FAULTS/ TECHNICAL EVENTS, NO AMBIENT STATED CONDITIONS AND NO DEVICE-RELATED DEVIATIONS. THE FIELD ENGINEER IN BERN DID NOT FOUND DEVIATIONS ON THE SERVICE SOFTWARE TEST. THE MANUFACTURER INTENDS TO CONTINUE THE INVESTIGATION BY OBTAINING ADDITIONAL INFORMATION. THE MANUFACTURER SECURED AND REVIEWED THE AVAILABLE LOG FILES. AS SOON AS NEW INFORMATION BECOMES AVAILABLE, WE WILL PROVIDE AN UPDATE; OTHERWISE, WE WILL REPORT BACK NO LATER THAN THE AGREED FOLLOW-UP DATE.
HAMILTON MEDIAL AG RECEIVED THE FOLLOWING COMPLAINT DESCRIPTION (SUMMARY): THE USER OBSERVED THAT THE ACTIVE HUMIDIFICATION ON THE VENTILATOR WAS SWITCHED OFF AND, AFTER REACTIVATION, THE BREATHING CIRCUIT FILLED WITH WATER, CAUSING LEAKAGE WHEN THE EXPIRATORY FILTER WAS DISCONNECTED. THE PATIENT WAS BRIEFLY DISCONNECTED, VENTILATED MANUALLY, AND THE VENTILATOR WAS REPLACED; THE PATIENT REMAINED STABLE. THE INVESTIGATION TO VERIFY THE ALLEGATION IS STILL IN PROGRESS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 113477 | HAMILTON-C6 | HAMILTON-C6 | CBK | HAMILTON MEDICAL AG | 160021 | 07630002808590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |