FDA Adverse Event Malfunction Summary report: N

HAMILTON-C6

MDR report key: 23836284 · Received December 18, 2025

Report

Report Number
3001421318-2025-00978
Event Type
Malfunction
Date Received
December 18, 2025
Date of Event
December 2, 2025
Report Date
May 20, 2026
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002808590
PMA / PMN Number
K201658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AGS REFERENCE: (B)(4). HAMILTON MEDICAL AGS CONCLUSION: INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AGS REFERENCE: (B)(4); HAMILTON MEDICAL AGS FOLLOW UP: IT WAS REPORTED THAT THE ACTIVE HUMIDIFICATION WAS FOUND SWITCHED OFF AND THAT THE BREATHING CIRCUIT FILLED WITH WATER AFTER REACTIVATION.THE LOGFILE ANALYSIS SHOWED NO TECHNICAL FAULTS/ TECHNICAL EVENTS, NO AMBIENT STATED CONDITIONS AND NO DEVICE-RELATED DEVIATIONS. THE FIELD ENGINEER IN BERN DID NOT FOUND DEVIATIONS ON THE SERVICE SOFTWARE TEST. THE MANUFACTURER INTENDS TO CONTINUE THE INVESTIGATION BY OBTAINING ADDITIONAL INFORMATION. THE MANUFACTURER SECURED AND REVIEWED THE AVAILABLE LOG FILES. AS SOON AS NEW INFORMATION BECOMES AVAILABLE, WE WILL PROVIDE AN UPDATE; OTHERWISE, WE WILL REPORT BACK NO LATER THAN THE AGREED FOLLOW-UP DATE.

Description of Event or Problem · 0

HAMILTON MEDIAL AG RECEIVED THE FOLLOWING COMPLAINT DESCRIPTION (SUMMARY): THE USER OBSERVED THAT THE ACTIVE HUMIDIFICATION ON THE VENTILATOR WAS SWITCHED OFF AND, AFTER REACTIVATION, THE BREATHING CIRCUIT FILLED WITH WATER, CAUSING LEAKAGE WHEN THE EXPIRATORY FILTER WAS DISCONNECTED. THE PATIENT WAS BRIEFLY DISCONNECTED, VENTILATED MANUALLY, AND THE VENTILATOR WAS REPLACED; THE PATIENT REMAINED STABLE. THE INVESTIGATION TO VERIFY THE ALLEGATION IS STILL IN PROGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113477 HAMILTON-C6 HAMILTON-C6 CBK HAMILTON MEDICAL AG 160021 07630002808590

Patients

Seq Age Sex Outcome Treatment
1