EXTENSION SET, 17 INCH, 0.2 MICRON FILTER, CLAVE(TM) Y-SITE, SECURE LOCK
Report
- Report Number
- 9615050-2025-00623
- Event Type
- Malfunction
- Date Received
- December 17, 2025
- Date of Event
- December 12, 2025
- Report Date
- January 21, 2026
- Manufacturer
- ICU MEDICAL COSTA RICA LTD.
- Product Code
- FPA
- UDI-DI
- 10887787005919
- PMA / PMN Number
- K151969
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DEVICE HAS BEEN REQUESTED FOR EVALUATION; HOWEVER, IT HAS NOT YET BEEN RECEIVED.
H10 - ADDITIONAL RELATED REPORT NUMBERS. 9615050-2025-00625-00. 9615050-2025-00624-00. SIXTEEN NEW EXTENSION SETS WERE RECEIVED FOR EVALUATION ON 12/29/2025. AS RECEIVED, NO DAMAGE OR ANOMALIES WERE NOTED. ONE PHOTO WAS RECEIVED FOR REVIEW ALSO AND IT SHOWED A SET WITH LEAKAGE AT THE INLINE FILTER. THE FILTER ON THE SIXTEEN NEW SETS WERE LEAK TESTED AND NO LEAKAGE WAS OBSERVED ON ANY OF THEM. THE REPORTED COMPLAINT OF LEAKAGE AT THE FILTER CANNOT BE REPLICATED BUT CAN BE CONFIRMED FOR ONE SET BASED ON THE CUSTOMER PROVIDED PHOTO. THE PROBABLE CAUSE CANNOT BE DETERMINED WITHOUT THE RETURN OF THE REPORTED SET. THE LOT HISTORY WAS REVIEWED, AND NO NONCONFORMITIES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.
THE EVENT INVOLVED AN EXTENSION SET, 17INCH, 0.2 MICRON FILTER, CLAVE (TM), Y-SITE, SECURE LOCK AND IT WAS REPORTED THAT THE DEVICE HAS A LEAKING ISSUE DURING INFUSION OF MIXED CHEMOTHERAPY (DOXORUBICIN, ETOPOSIDE AND VINCRISTINE). THERE WAS PATIENT INVOLVEMENT, NO HARM AND DID CAUSE A DELAY IN THERAPY. THIS REPORT CAPTURES THIRD OF FOUR INCIDENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2230448 | EXTENSION SET, 17 INCH, 0.2 MICRON FILTER, CLAVE(TM) Y-SITE, SECURE LOCK | SET, ADMINISTRATION, INTRAVASCULAR | FPA | ICU MEDICAL COSTA RICA LTD. | 14338083 | 10887787005919 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DOXORUBICIN, UNK MFR.| ETOPOSIDE, UNK MFR.| VINCRISTINE, UNK MFR. |