FDR VISIONARY SUITE B40 UL 80KW GENERATOR SA42
Report
- Report Number
- 1000513161-2025-00034
- Event Type
- Injury
- Date Received
- December 16, 2025
- Date of Event
- December 4, 2025
- Report Date
- February 19, 2026
- Manufacturer
- SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
- Product Code
- KPR
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PER AGREEMENT WITH CUSTOMER, THE CUSTOMER IS RESPONSIBLE TO DESIGN AND IMPLEMENT AN ASSEMBLY THAT MEETS THE SYSTEM'S REQUIREMENTS FOR THE SPECIFIC ROOM WHERE THE EQUIPMENT WILL BE DEPLOYED. CUSTOMER PROVIDED A TIMELINE OF DEMOLITION OF ROOM AND STRUCTURAL REPAIR AS PART OF THE INSTALL OF THE NEW EQUIPMENT. EXPECTED COMPLETION OF CONSTRUCTION IS END OF FEBRUARY 2026 WITH INSTALLATION TO FOLLOW. INITIAL IMPORTER MDR 1000513161-2025-00034 WAS SUBMITTED DUE TO SAFETY OF THE TECH. REF: FUJIFILM HEALTHCARE AMERICAS CORPORATION COMPLAINT #(B)(4) (FOR TECH)
ON (B)(6) 2025, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING THE FDR VISIONARY SUITE B40 UL 80KW GENERATOR SA42. IT WAS REPORTED THAT THE OTCX FELL DOWN FROM THE CEILING IN VISIONARY SUITE ROOM 1 AT THE FACILITY. IT BROKE FROM THE CEILING AND COLLAPSED, AT LEAST IN PART, ON THE PATIENT WHO WAS BEING IMAGED. THE TECH WHO WAS COMPLETING THE IMAGING WAS BLOWN BACK AND HIT THE FLOOR. TECH WENT TO SEE PRIMARY CARE DOCTOR AND IS NOW OUT OF WORK FOR THE WEEK. THE ROOM HAS BEEN SHUT DOWN, AND NO PATIENTS ARE BEING SCANNED AT THIS TIME. DUE TO SAFETY OF THE TECH, FUJIFILM HEALTHCARE AMERICAS CORPORATION IS REPORTING THIS EVENT. REF: FUJIFILM HEALTHCARE AMERICAS CORPORATION COMPLAINT #(B)(4) (FOR TECH)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1875145 | FDR VISIONARY SUITE B40 UL 80KW GENERATOR SA42 | SYSTEM, X-RAY, STATIONARY | KPR | SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION | FDR VISIONARY SUITE B40 UL 80KW GENERATOR SA42 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |