FDA Adverse Event Malfunction Summary report: N

CARDIOMEMS HOSPITAL ELECTRONICS SYSTEM

MDR report key: 23817692 · Received December 16, 2025

Report

Report Number
3004936110-2025-06597
Event Type
Malfunction
Date Received
December 16, 2025
Date of Event
October 7, 2025
Report Date
December 16, 2025
Manufacturer
ABBOTT MEDICAL
Product Code
MOM
UDI-DI
05415067029331
PMA / PMN Number
P100045
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ONE CARDIOMEMS HOSPITAL ELECTRONICS SYSTEM WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION OF RETURNED MATERIAL REVEALED FUNCTIONAL DAMAGE TO THE GLOBAL SYSTEM FOR MOBILE COMMUNICATIONS (GSM) ANTENNA RESULTING IN EXPOSED CIRCUITRY/WIRES. DAMAGE TO THE GSM ANTENNA CAME IN THE FORM OF THE SUBMINIATURE VERSION A (SMA) CONNECTOR PORTION BROKEN OFF FROM THE REST OF THE COMPONENT. ONLY THE BROKEN OFF SMA CONNECTOR PORTION OF THE GSM ANTENNA WAS RETURNED. FUNCTIONAL DAMAGE WAS ALSO FOUND ON THE EXTERNAL PRINTER SERIAL CABLE ASSEMBLY IN THE FORM OF THE CONNECTOR D-SUB HOUSING RECEPTACLE BROKEN OFF FROM THE REST OF THE COMPONENT, RESULTING IN EXPOSED CIRCUITRY/WIRES. ONLY THE BROKEN OFF CONNECTOR D-SUB HOUSING RECEPTACLE OF THE PRINTER SERIAL CABLE WAS RETURNED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.

Description of Event or Problem · 0

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS, IN WHICH THE GLOBAL SYSTEM FOR MOBILE COMMUNICATIONS (GSM) ANTENNA AND PRINTER SERIAL CABLE ASSEMBLY WERE FOUND TO BE BROKEN WITH EXPOSED CIRCUITRY/WIRES. THE HOSPITAL SYSTEM WAS REPLACED AND THERE WERE NO ADVERSE CONSEQUENCES REPORTED, AND NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2444704 CARDIOMEMS HOSPITAL ELECTRONICS SYSTEM System, hemodynamic, implantable MOM ABBOTT MEDICAL CM3000 7148153 05415067029331

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown