FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SERUM

MDR report key: 23816937 · Received December 16, 2025

Report

Report Number
1024879-2025-02128
Event Type
Malfunction
Date Received
December 16, 2025
Date of Event
November 15, 2025
Report Date
March 10, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903678205
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: G.4. PMA / 510(K)#: K960250. INVESTIGATION SUMMARY: BD RECEIVED 4 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND UNDERFILL WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: UNDERFILL. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. BASED ON THE LOW DEFECT RATE FOR THE BATCH IN QUESTION, NO ACTIONS ARE PLANNED AT THIS TIME COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, SIX TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® SERUM, SIX TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277933 BD VACUTAINER® SERUM TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 5182823 30382903678205

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown