STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-21910
- Event Type
- Injury
- Date Received
- December 16, 2025
- Date of Event
- June 6, 2025
- Report Date
- February 26, 2026
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628001503
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON FEB 13, 2026 WITH LOT NUMBER 2183647. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ RUPTURE: OBSERVED BROKEN DEVICE THROUGH MICROSCOPIC INSPECTION ASSESSED AS FOLD FLAW OPENING AND OBSERVED A MISSING PIECE OF SHELL THROUGH MICROSCOPIC INSPECTION ASSESSED AS INCONCLUSIVE. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (NON-PENETRATING NICKS) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, H6.
HEALTHCARE PROFESSIONAL REPORTED RUPTURE. LATER, HEALTHCARE PROFESSIONAL REPORTED "BLOODY NIPPLE DISCHARGE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED RUPTURE. THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.
DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2281718 | STYLE 15 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2183647 | 10888628001503 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Female | Required Intervention |