FDA Adverse Event Injury Summary report: N

QUICK SET

MDR report key: 23813813 · Received December 16, 2025

Report

Report Number
3003442380-2025-17249
Event Type
Injury
Date Received
December 16, 2025
Date of Event
September 26, 2025
Report Date
February 23, 2026
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244017573
PMA / PMN Number
K160648
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 06-FEB-2026 AGAINST "LOT NUMBER 6013472 AND SIMILAR MALFUNCTION CODES: NO MALFUNCTION BASED ON COMPLAINT INFORMATION. NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE REVIEW CONFIRMED THAT LOT 6013472 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NONCONFORMING REPORTS (NCR)S OR CORRECTIVE AND PREVENTIVE ACTION (CAPA)S OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 06-FEB-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL 6013472 AND SIMILAR MALFUNCTION CODES NO MALFUNCTION BASED ON COMPLAINT INFORMATION. NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE COUNT OF COMPLAINT IS 3. THE COMPLAINT NUMBERS ARE (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6013472 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND MANUFACTURED IN THE MULTIVAC 12, ON 26-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. RETAIN SAMPLES TESTING: BASED ON THE CLASSIFICATION OF THE MALFUNCTION CODE, THIS TYPE OF ISSUE DOES NOT REQUIRE VISUAL TESTING, FUNCTIONAL TESTING, OR EVALUATION OF RETAIN SAMPLES, AS IT CORRESPONDS TO CATEGORIES SUCH AS "MISUSE, USER RELATED ISSUES, OR NO MALFUNCTION ALLEGED." THEREFORE, NO RETAIN SAMPLES WERE REQUESTED AND NO PRODUCT TESTING WAS PERFORMED. CONCLUSION: NO FURTHER INVESTIGATION ACTIVITIES WERE REQUIRED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6013472 AND RELATED MALFUNCTION CODES. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. BASED ON THE CLASSIFICATION OF THE MALFUNCTION CODE, THIS TYPE OF ISSUE DOES NOT REQUIRE VISUAL TESTING, FUNCTIONAL TESTING, OR EVALUATION OF RETAIN SAMPLES, AS IT CORRESPONDS TO CATEGORIES SUCH AS "MISUSE, USER RELATED ISSUES, OR NO MALFUNCTION ALLEGED." THEREFORE, NO RETAIN SAMPLES WERE REQUESTED AND NO PRODUCT TESTING WAS PERFORMED.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED HIGH BLOOD GLUCOSE (BG) EVENT LEADING TO HOSPITALIZATION ON (B)(6) 2025. THE PATIENT BLOOD GLUCOSE (BG) WAS 1427 MG/DL AND WAS UNCONSCIOUS. THE PATIENT HAD HIGH KETONES. THE PATIENT GOT TREATED WITH INSULIN VIA MANUAL INJECTION. THE LENGTH OF HOSPITALIZATION WAS GREATER THAN 24 HOURS. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620946 QUICK SET UNO QUICK-SET 60/9 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-397A 6013472 05705244017573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H