FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE VALVE

MDR report key: 2381237 · Received December 19, 2011

Report

Report Number
1226348-2011-00494
Event Type
Injury
Date Received
December 19, 2011
Date of Event
November 12, 2011
Manufacturer
CODMAN AND SHURTLEFF - MEDOS
Product Code
JXG
PMA / PMN Number
K980778
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AK, GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PRODUCT CODE IS ACTUALLY THAT OF 82-3114 AND NOT THAT OF 82-3112 AS REPORTED IN THE INITIAL MEDWATCH REPORT. EXAMINATION OF THE VALVE DETERMINED THAT BIOLOGICAL DEBRIS WITHIN THE DEVICE LIKELY CAUSED THE DIFFICULTY EXPERIENCED BY THE CUSTOMER. THIS BIOLOGICAL DEBRIS WAS DISLODGED DURING FLOW TESTING OF THE RETURNED VALVE. AFTER DISLODGING, THE VALVE PASSED THE PRESSURE TESTS BUT NOT THE PROGRAMMING. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE VALVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.

Description of Event or Problem · 1

VALVE COULD NOT BE REPROGRAMMED AND AS A RESULT IT WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAKIM PROGRAMMABLE VALVE CNS SHUNT JXG CODMAN AND SHURTLEFF - MEDOS CLKB0V

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention