HAKIM PROGRAMMABLE VALVE
Report
- Report Number
- 1226348-2011-00494
- Event Type
- Injury
- Date Received
- December 19, 2011
- Date of Event
- November 12, 2011
- Manufacturer
- CODMAN AND SHURTLEFF - MEDOS
- Product Code
- JXG
- PMA / PMN Number
- K980778
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, GM
- Reporter Occupation
- OTHER
Narratives
HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.
THE PRODUCT CODE IS ACTUALLY THAT OF 82-3114 AND NOT THAT OF 82-3112 AS REPORTED IN THE INITIAL MEDWATCH REPORT. EXAMINATION OF THE VALVE DETERMINED THAT BIOLOGICAL DEBRIS WITHIN THE DEVICE LIKELY CAUSED THE DIFFICULTY EXPERIENCED BY THE CUSTOMER. THIS BIOLOGICAL DEBRIS WAS DISLODGED DURING FLOW TESTING OF THE RETURNED VALVE. AFTER DISLODGING, THE VALVE PASSED THE PRESSURE TESTS BUT NOT THE PROGRAMMING. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE VALVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.
VALVE COULD NOT BE REPROGRAMMED AND AS A RESULT IT WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HAKIM PROGRAMMABLE VALVE | CNS SHUNT | JXG | CODMAN AND SHURTLEFF - MEDOS | CLKB0V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |