FDA Adverse Event Injury Summary report: N

GALAXY SYSTEM

MDR report key: 23804016 · Received December 15, 2025

Report

Report Number
3021325287-2025-00069
Event Type
Injury
Date Received
December 15, 2025
Date of Event
September 11, 2024
Report Date
December 12, 2025
Manufacturer
NOAH MEDICAL CORP
Product Code
EOQ
PMA / PMN Number
K223144
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE SCOPE WAS NOT RETURNED FOR INVESTIGATION, BUT MANUFACTURING RECORDS CONFIRMED IT PASSED FINAL QA/QC. VIDEO REVIEW IDENTIFIED NO SYSTEM MALFUNCTIONS OR USE ERRORS. THE PHYSICIAN DID NOT ATTRIBUTE THE BLEEDING TO THE GALAXY SYSTEM, STATING IT WAS RELATED TO BIOPSY ACTIVITY. VIDEO REVIEW SUPPORTS THIS CONCLUSION, AS BLEEDING ONSET AND ESCALATION CORRESPONDED DIRECTLY WITH SAMPLING AND NO DEVICE-RELATED ISSUES WERE IDENTIFIED.

Description of Event or Problem · 0

AFTER COMPLETING SAMPLING OF A RIGHT UPPER LOBE (RUL) LESION, THE PHYSICIAN BEGAN WITHDRAWING THE GALAXY BRONCHOSCOPE BUT WAS UNABLE TO VISUALIZE THE AIRWAY BECAUSE THE CAMERA VIEW HAD BECOME COMPLETELY RED. A FLUOROSCOPY IMAGE CONFIRMED THE SCOPE WAS ALREADY NEAR THE TRACHEA. THE GALAXY SCOPE WAS RAPIDLY REMOVED, AND A THERAPEUTIC SCOPE WAS CALLED FOR. WHILE THE THERAPEUTIC SCOPE WAS BEING PREPARED, BLOOD WAS OBSERVED RISING INTO THE ENDOTRACHEAL TUBE (ETT). THE PHYSICIAN IMMEDIATELY BEGAN SUCTIONING AND WAS ABLE TO RE-ENTER THE RUL, WEDGING THE BRONCHOSCOPE TO TAMPONADE THE BLEEDING. THE PATIENT WAS PLACED IN THE RIGHT-SIDE-DOWN POSITION, ICE-COLD SALINE WAS INSTILLED MULTIPLE TIMES, AND A FOGARTY BALLOON WAS ALSO USED FOR TAMPONADE. AFTER SEVERAL MINUTES, BLEEDING STOPPED, AND THE AIRWAYS WERE SUCTIONED CLEAR. THE PATIENT WAS EXTUBATED, TAKEN TO PACU IN STABLE CONDITION, RECEIVED A CHEST X-RAY, AND WAS LATER DISCHARGED HOME. ATTEMPTS TO GATHER ADDITIONAL PATIENT DEMOGRAPHICS WERE UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
789523 GALAXY SYSTEM GALAXY SYSTEM BRONCHOSCOPE GALB-001 EOQ NOAH MEDICAL CORP GALB-001 2024072401

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Required Intervention