INSET II
Report
- Report Number
- 3003442380-2025-17116
- Event Type
- Injury
- Date Received
- December 15, 2025
- Date of Event
- November 3, 2025
- Report Date
- November 28, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018129
- PMA / PMN Number
- K032854
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-17116. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6013144, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 27-NOV-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL 6013144. THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6013144 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 123 AND MANUFACTURED IN THE INSET LINE L1 L115 ON 08-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE ASSEMBLY, LOT 5C06115 WAS MANUFACTURED ACCORDING TO THE WI VERSION 9 AND MANUFACTURED IN THE ITL01, ON 23-APR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED. IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE(S) FROM THE LOT HAVE BEEN REQUESTED. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH: WI GUIDANCE FOR VISUAL TEST FOR COMPLAINTS AREA VERSION 3: ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. WI GUIDANCE FOR FUNCTIONAL TESTING 1 AIR FLOW TEST FOR COMPLAINTS AREA VERSION 2: ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. WI GUIDANCE FOR FUNCTIONAL TESTING 2 AIR LEAK TEST FOR COMPLAINTS AREA VERSION 2: ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, 1 COMPLAINT IN THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
INITIAL AND FINAL (B)(4) - DEVICE 5 OF 8.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INFECTION AT 8 INFUSION SETS INSERTION SITE FROM (B)(6) 2025. THE INFUSION SET WAS IN USE FOR 1 DAY. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 325309 | INSET II | UNO INSET II 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002817 | 6013144 | 05705244018129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Male |