FDA Adverse Event
Malfunction
Summary report: N
HAMILTON-C6
MDR report key: 23797517
·
Received December 13, 2025
Report
- Report Number
- 3001421318-2025-00971
- Event Type
- Malfunction
- Date Received
- December 13, 2025
- Date of Event
- November 20, 2025
- Report Date
- December 13, 2025
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002808590
- PMA / PMN Number
- K201658
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
HAMILTON MEDICAL AG REF. NR: (B)(4); HAMILTON MEDICAL AGS CONCLUSION: THE ON-SITE HAMILTON MEDICAL INC. SERVICE TECHNICIAN CONFIRMED THAT THE DEVICE DISPLAY WENT BLACK WHILE CONNECTED TO A PATIENT. LOG FILE ANALYSIS VERIFIED THE EVENT, SHOWING THE ENTRIES ¿PANEL CONNECTION LOST¿ AND ¿PANEL RECONNECTED.¿ TO ADDRESS THE ISSUE, THE TECHNICIAN REPORTED THAT BOTH THE PROCESSOR BOARD AND THE VU ESM BOARD ARE BEING REPLACED.
Description of Event or Problem · 0
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: CUSTOMER REPORTS DISPLAY WENT BLACK WHILE ON A PATIENT. AN ADDITIONAL DEVICE WAS REQUIRED. NO FURTHER CLINICAL SIGNS, SYMPTOMS OR CONDITIONS AND NO HEALTH CONSEQUENCES OR IMPACT, WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2762227 | HAMILTON-C6 | HAMILTON-C6 | CBK | HAMILTON MEDICAL AG | 160021 | 07630002808590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |