FDA Adverse Event Malfunction Summary report: N

HAMILTON-C6

MDR report key: 23797517 · Received December 13, 2025

Report

Report Number
3001421318-2025-00971
Event Type
Malfunction
Date Received
December 13, 2025
Date of Event
November 20, 2025
Report Date
December 13, 2025
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002808590
PMA / PMN Number
K201658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG REF. NR: (B)(4); HAMILTON MEDICAL AGS CONCLUSION: THE ON-SITE HAMILTON MEDICAL INC. SERVICE TECHNICIAN CONFIRMED THAT THE DEVICE DISPLAY WENT BLACK WHILE CONNECTED TO A PATIENT. LOG FILE ANALYSIS VERIFIED THE EVENT, SHOWING THE ENTRIES ¿PANEL CONNECTION LOST¿ AND ¿PANEL RECONNECTED.¿ TO ADDRESS THE ISSUE, THE TECHNICIAN REPORTED THAT BOTH THE PROCESSOR BOARD AND THE VU ESM BOARD ARE BEING REPLACED.

Description of Event or Problem · 0

HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: CUSTOMER REPORTS DISPLAY WENT BLACK WHILE ON A PATIENT. AN ADDITIONAL DEVICE WAS REQUIRED. NO FURTHER CLINICAL SIGNS, SYMPTOMS OR CONDITIONS AND NO HEALTH CONSEQUENCES OR IMPACT, WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2762227 HAMILTON-C6 HAMILTON-C6 CBK HAMILTON MEDICAL AG 160021 07630002808590

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown