FDA Adverse Event Malfunction Summary report: N

BD ALARIS PUMP MODULE SMARTSITE INFUSION SET

MDR report key: 23796420 · Received December 12, 2025

Report

Report Number
9616066-2025-03767
Event Type
Malfunction
Date Received
December 12, 2025
Date of Event
October 27, 2025
Report Date
January 29, 2026
Manufacturer
SISTEMAS MEDICOS ALARIS S.A DE C.V.
Product Code
FPA
UDI-DI
07613203021012
PMA / PMN Number
K221327
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

ONE SAMPLE MODEL 2420-0007 LOT 25075647 WAS RETURNED FOR INVESTIGATION. THE SET WAS EXAMINED FOR DEFECTS AND ABNORMALITIES. FOREIGN MATTER WAS SEEN ON THE INSIDE OF THE SPIKE CAP. NO OTHER DEFECTS OR ABNORMALITIES WERE OBSERVED. A QUALITY NOTIFICATION WAS SENT TO THE SUPPLIER. NO RESPONSE WAS RECEIVED BY THE SUPPLIER. IF A RESPONSE IS RECEIVED. THE INVESTIGATION WILL BE REOPENED TO INCLUDE THE INFORMATION THEY PROVIDE. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED. THERE WERE NO QUALITY NOTIFICATIONS ISSUED FOR THE FAILURE MODE REPORTED BY THE CUSTOMER DURING THE PRODUCTION BUILD OF THIS SET.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ALARIS PUMP MODULE SMARTSITE INFUSION SET HAD FOREIGN MATTER IN THE FLUID PATHWAY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONTAMINANT LOCATED IN AN AREA THAT SHOULD BE STERILE AS INTRAVENOUS FLUIDS WILL BE ACCESSED USING THIS SPIKED END OF THE INFUSION TUBING. DESCRIPTION OF PRODUCT: GEM V/NV 20D 1CV 2SS DEHP-FREE 20EA/CA. LOT OR S/N: (B)(6). COMPLAINT NOTICED: PRIOR TO USE. PROBLEM FREQUENCY: 1ST TIME. PATIENT INJURY: NO. QTY AFFECTED: 1 EA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2769851 BD ALARIS PUMP MODULE SMARTSITE INFUSION SET INTRAVASCULAR ADMINISTRATION SET FPA SISTEMAS MEDICOS ALARIS S.A DE C.V. 25075647 07613203021012

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown