FDA Adverse Event Malfunction Summary report: N

VANTAGE

MDR report key: 237890 · Received August 27, 1999

Report

Report Number
2916556-1999-00179
Event Type
Malfunction
Date Received
August 27, 1999
Date of Event
April 8, 1999
Report Date
April 8, 1999
Manufacturer
ADAC LABORATORIES
Product Code
IYX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED MCD SINGLES COUNT RATE IS NOT DOCUMENTED IN THE MCD USERS MANUAL. CURRENTLY, ADAC SUGGESTS TO THE CUSTOMER TO KEEP THE SINGLES COUNT RATE BETWEEN 800-1600KCPS. IT HAS BEEN SEEN THAT IF THE SINGLES COUNT RATE EXCEEDS THESE VALUES, THERE COULD BE ARTIFACTS IN THE PROCESSED DATA FOR MCD STUDIES. THERE WAS ALSO AN ASSOCIATED REPORT OF SYSTEM LOCK-UP AT THE SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VANTAGE NUCLEAR MEDICINE EQUIPMENT IYX ADAC LABORATORIES 2146-3260A NA

Patients

Seq Age Sex Outcome Treatment
1 *