FDA Adverse Event
Malfunction
Summary report: N
VANTAGE
MDR report key: 237890
·
Received August 27, 1999
Report
- Report Number
- 2916556-1999-00179
- Event Type
- Malfunction
- Date Received
- August 27, 1999
- Date of Event
- April 8, 1999
- Report Date
- April 8, 1999
- Manufacturer
- ADAC LABORATORIES
- Product Code
- IYX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED MCD SINGLES COUNT RATE IS NOT DOCUMENTED IN THE MCD USERS MANUAL. CURRENTLY, ADAC SUGGESTS TO THE CUSTOMER TO KEEP THE SINGLES COUNT RATE BETWEEN 800-1600KCPS. IT HAS BEEN SEEN THAT IF THE SINGLES COUNT RATE EXCEEDS THESE VALUES, THERE COULD BE ARTIFACTS IN THE PROCESSED DATA FOR MCD STUDIES. THERE WAS ALSO AN ASSOCIATED REPORT OF SYSTEM LOCK-UP AT THE SITE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VANTAGE | NUCLEAR MEDICINE EQUIPMENT | IYX | ADAC LABORATORIES | 2146-3260A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |