IMP,TSV,4.7,10,MTX,MG
Report
- Report Number
- 0001038806-2025-03791
- Event Type
- Malfunction
- Date Received
- December 11, 2025
- Report Date
- February 2, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- UDI-DI
- 00889024019997
- PMA / PMN Number
- K101977
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODE WAS ADDED: 10, 4109, 3331 AND 4111. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 180. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4307. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED THE REPORTED DEVICE FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICES IDENTIFIED THE IMPLANT, AND BUNDLE MOUNT WITH SIGNS OF USE, DEBRIS, AND MINOR APPARENT BONE / TISSUE ATTACHED TO THE IMPLANT'S EXTERNAL THREADS. THE IMPLANT AND MOUNT WERE OBSERVED ATTACHED TOGETHER, AND DURING A PHYSICAL / FUNCTIONAL TEST THE IMPLANT AND MOUNT DID NOT DISENGAGE/RELEASE AS INTENDED. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION IS INCORRECT TECHNIQUE USED DURING IMPLANT PLACEMENT - CUSTOMER ERROR. THEREFORE, BASED ON THE AVAILABLE INFORMATION AND FUNCTIONAL TESTING, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
ZIMVIE COMPLAINT NUMBER (B)(4). G4: ADDITIONAL PMA/510(K) NUMBER ¿ K101880. PRODUCT HAS BEEN RECEIVED BY ZIMVIE AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT THE IMPLANT ANALOG AND OR CARRIER WAS INITIALLY LOOSENED FROM THE IMPLANT PRIOR TO PLACEMENT, HOWEVER IT BECAME FUSED DURING PLACEMENT. IT WAS CONFIRMED THAT ANOTHER IMPLANT WAS USED TO COMPLETE THE PROCEDURE. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2413720 | IMP,TSV,4.7,10,MTX,MG | DENTAL IMPLANT | DZE | ZIMVIE US CORP LLC | 1289716 | 00889024019997 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male |