FDA Adverse Event Injury Summary report: N

WAVEWRITER ALPHA? 16

MDR report key: 23782011 · Received December 11, 2025

Report

Report Number
3006630150-2025-11353
Event Type
Injury
Date Received
December 11, 2025
Date of Event
May 13, 2025
Report Date
January 19, 2026
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729985082
PMA / PMN Number
P030017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218700. MODEL: SC-2218-70. SERIAL: (B)(6). BATCH: 240374. UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218700. MODEL: SC-2218-70. SERIAL: (B)(6). BATCH: 240512. UDI: (B)(4).

Additional Manufacturer Narrative · 0

B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218700. MODEL: SC-2218-70. SERIAL: (B)(6). BATCH: (B)(6). UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218700. MODEL: SC-2218-70. SERIAL: (B)(6). BATCH: (B)(6). UDI: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THESE SPINAL CORD STIMULATION (SCS) DEVICES WERE REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICES IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THESE SPINAL CORD STIMULATION (SCS) DEVICES WERE REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICES IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE SCS PATIENT UNDERWENT THE REPLACEMENT PROCEDURE OF THE IMPLANTABLE PULSE GENERATOR (IPG) AND LEADS DUE TO HIGH IMPEDANCES. IT WAS FURTHER NOTED THAT, ASIDE FROM THE PRIOR REASON, THE REPLACEMENT WAS ALSO DUE TO MAGNETIC RESONANCE IMAGING (MRI) REQUIREMENTS. THE IPG AND LEADS WERE DISPOSED PER FACILITY PROTOCOL AND WILL NOT BE RETURNED FOR ANALYSIS. POSTOPERATIVELY, THE PATIENT CONDITION WAS NORMAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2553955 WAVEWRITER ALPHA? 16 STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-1216 506935 08714729985082

Patients

Seq Age Sex Outcome Treatment
1 57 YR Male Required Intervention