FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE

MDR report key: 23768656 · Received December 10, 2025

Report

Report Number
9610773-2025-08089
Event Type
Malfunction
Date Received
December 10, 2025
Date of Event
November 12, 2025
Report Date
December 10, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE OF THE BROKEN R-UNIT WAS TRACED TO COMPONENT FAILURE. THE MOST PROBABLE CAUSE OF THE FOREIGN MATERIAL IN THE LLK PLUG OF THE VIDEO CABLE SECTION WAS NOT ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GYNECOLOGIC LAPAROSCOPE EXHIBITED THE CHARGE COUPLED DEVICE (R-UNIT) WAS BROKEN, AND THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION CONTAINS FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2166903 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC HET OLYMPUS WINTER & IBE GMBH WA50042A 04042761074971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown