FDA Adverse Event
Malfunction
Summary report: N
VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE
MDR report key: 23768656
·
Received December 10, 2025
Report
- Report Number
- 9610773-2025-08089
- Event Type
- Malfunction
- Date Received
- December 10, 2025
- Date of Event
- November 12, 2025
- Report Date
- December 10, 2025
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HET
- UDI-DI
- 04042761074971
- PMA / PMN Number
- K190744
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE OF THE BROKEN R-UNIT WAS TRACED TO COMPONENT FAILURE. THE MOST PROBABLE CAUSE OF THE FOREIGN MATERIAL IN THE LLK PLUG OF THE VIDEO CABLE SECTION WAS NOT ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem · 0
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GYNECOLOGIC LAPAROSCOPE EXHIBITED THE CHARGE COUPLED DEVICE (R-UNIT) WAS BROKEN, AND THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION CONTAINS FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2166903 | VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE | LAPAROSCOPE, GYNECOLOGIC | HET | OLYMPUS WINTER & IBE GMBH | WA50042A | 04042761074971 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |