FDA Adverse Event Malfunction Summary report: N

SECHRIST INDUSTRIES INC.

MDR report key: 23759634 · Received December 9, 2025

Report

Report Number
2020676-2025-00030
Event Type
Malfunction
Date Received
December 9, 2025
Date of Event
November 12, 2025
Report Date
December 9, 2025
Manufacturer
SECHRIST INDUSTRIES
Product Code
DTX
UDI-DI
00899660002345
PMA / PMN Number
K023745
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE OF UNABLE TO OXYGENATE WAS CONFIRMED. THE RETURNED MIXER HAD PREVIOUSLY BEEN SERVICED BY A THIRD PARTY. DURING EVALUATION, THE FIO2 KNOB WAS FOUND TO BE LOOSE AND DETACHED, AND THE BALANCE WAS OUTSIDE THE SPECIFIED LIMITS. THE MOST LIKELY ROOT CAUSE IS INADEQUATE SERVICING OF THE DEVICE. AT THIS TIME THERE IS NO EVIDENCE THAT A MANUFACTURING NON-CONFORMITY CONTRIBUTED TO THE REPORTED COMPLAINT, AND THE INSTRUCTIONS FOR USE WERE REVIEWED AND DETERMINED TO PROVIDE ADEQUATE INSTRUCTIONS AND WARNINGS FOR THE SAFE AND EFFECTIVE USE OF THE DEVICE. THE IFU STATES, "TO ENSURE PROPER FUNCTION AND ACCURACY, THE SECHRIST AIR/OXYGEN GAS MIXERS MUST BE THOROUGHLY OVERHAULED EVERY TWO (2) YEARS. TO MAINTAIN THE PRODUCT WARRANTY, THE OVERHAUL MUST BE PERFORMED BY SECHRIST INDUSTRIES OR BY SECHRIST AUTHORIZED PERSONNEL." THEREFORE, NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE NECESSARY. ALL COMPLAINTS ARE TRENDED AND REVIEWED BY MANAGEMENT ON A MONTHLY BASIS. AS PART OF THIS MONTHLY REVIEW, ANY EXCURSION ABOVE THE CONTROL LIMITS FOR THIS FAILURE MODE WILL BE ASSESSED, DOCUMENTED AND ACTED UPON AS WARRANTED. MANUFACTURER REFERENCE NO. (B)(4).

Description of Event or Problem · 0

CUSTOMER REPORTED STAFF WAS ATTEMPTING TO OXYGENATE THE PATIENT WHILE HE/SHE WAS ON BYPASS AND WERE UNABLE TO. STAFF SWAPPED THE DEVICE FOR ANOTHER ONE AND IT WORKED APPROPRIATELY. NO PATIENT HARM OCCURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2720818 SECHRIST INDUSTRIES INC. GAS CONTROL UNIT, CARDIOPULMONARY BYPASS DTX SECHRIST INDUSTRIES 3500CP-G 00899660002345

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown