MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2025-13630
- Event Type
- Injury
- Date Received
- December 8, 2025
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000549
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON 27-DEC-2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON 03-JAN-2026, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED A RUPTURE IN THE BREAST IMPLANT. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE BREAST IMPLANT WAS FOUND TO BE RUPTURED. IN ADDITION, AN AREA OF SHELL ABRASION WAS NOTED ON THE EDGES OF THE RUPTURE. THESE FINDINGS ARE CONSISTENT WITH NORMAL WEAR OF THE DEVICE. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE, WHICH MAY BE CAUSED BY CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE, SUCH AS AN EXCESSIVELY SMALL BREAST POCKET OR THE FOLDING OR WRINKLING OF THE SHELL WITHIN THE BREAST POCKET. IN SOME CASES, BREAST IMPLANTS MAY ALSO EXPERIENCE NORMAL WEAR OVER TIME. THE CONTRALATERAL DEVICE WAS ALSO RECEIVED (LOT NUMBER 6830514). THE PATIENT DID NOT REPORT ANY ISSUE WITH THIS DEVICE. THEREFORE, NO FURTHER ANALYSIS OF THE CONTRALATERAL DEVICE IS REQUIRED. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA, AND THAT NORMAL WEAR MAY HAVE CAUSED THE OBSERVED DEFECT. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: BREAST PROSTHESIS RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR MEMORYGEL BREAST IMPLANT 550CC GEL BREAST PROSTHESIS THAT RUPTURED POST IMPLANTATION. RUPTURE OF THE PATIENT¿S LEFT BREAST PROSTHESIS WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL BREAST IMPLANT 475CC GEL BREAST PROSTHESES ON (B)(6) 2025.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2007760 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 6830514 | 00081317000549 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | Required Intervention |