FDA Adverse Event Malfunction Summary report: N

RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS

MDR report key: 23741952 · Received December 8, 2025

Report

Report Number
9610773-2025-07843
Event Type
Malfunction
Date Received
December 8, 2025
Date of Event
November 14, 2025
Report Date
December 23, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HIH
PMA / PMN Number
K931995
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR CORRECTION TO D4 - LOT NUMBER (ADDED) AND NO SERIAL NUMBER.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D4, D10, E1 NAME AND ADDRESS, H2, H3, H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS), H11. CORRECTED FIELDS: G4 IS K931995, H5 SHOULD BE NO. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS CONFIRMED. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: COMPONENT FAILURE OF THE CERAMIC HEAD; ISSUE RELATED TO STRESS PROBLEM OR EXCESSIVE FORCE. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THE RESECTION SHEATH HEAD END WAS FRACTURED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142486 RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS A42011A HIH OLYMPUS WINTER & IBE GMBH A42011A 22808

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown