FDA Adverse Event Injury Summary report: N

LINEAR? ST

MDR report key: 23741210 · Received December 8, 2025

Report

Report Number
3006630150-2025-11243
Event Type
Injury
Date Received
December 8, 2025
Date of Event
July 9, 2025
Report Date
December 5, 2025
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729767749
PMA / PMN Number
P030017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BLOCK B3: EXACT DATE UNKNOWN, EVENT OCCURRED TWO MONTHS PRIOR TO THE DATE THE MANUFACTURER BECAME AWARE. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218700. MODEL: SC-2218-70. SERIAL: (B)(6). BATCH: 7093349. UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218500. MODEL: SC-2218-50. SERIAL: (B)(6). BATCH: 7093349. UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2218500. MODEL: SC-2218-50. SERIAL: (B)(6). BATCH: 7137135. UDI: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WERE SEVERAL IMPEDANCES ON ONE OF THE LEADS AND THE PATIENT EXPERIENCED LESS PAIN RELIEF SINCE. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE NOT RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156288 LINEAR? ST STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-2218-70 7095593 08714729767749

Patients

Seq Age Sex Outcome Treatment
1 69 YR Female Required Intervention