MDR report key: 2373968 · Received December 14, 2011

Report

Report Number
2649622-2011-18294
Event Type
Malfunction
Date Received
December 14, 2011
Report Date
August 31, 2021
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. SENSING - INTERFERENCE/NOISE VENTRICULAR SHORT INTERVAL COUNT V-SIC=47.6 COUNTS AVG/DAY, IN 450.71 DAYS, BETWEEN 19-JUL-2010 17:55:58 AND 13-OCT-2011 10:52:06.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD IS OVERSENSING. THE LEAD IS STILL IN USE. THE LEAD WILL BE MONITORED CLOSELY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) LWS MEDTRONIC PUERTO RICO, INC. 6945 ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Other (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB