FDA Adverse Event
Malfunction
Summary report: N
SPRINT
MDR report key: 2373968
·
Received December 14, 2011
Report
- Report Number
- 2649622-2011-18294
- Event Type
- Malfunction
- Date Received
- December 14, 2011
- Report Date
- August 31, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S13
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. SENSING - INTERFERENCE/NOISE VENTRICULAR SHORT INTERVAL COUNT V-SIC=47.6 COUNTS AVG/DAY, IN 450.71 DAYS, BETWEEN 19-JUL-2010 17:55:58 AND 13-OCT-2011 10:52:06.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD IS OVERSENSING. THE LEAD IS STILL IN USE. THE LEAD WILL BE MONITORED CLOSELY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) | LWS | MEDTRONIC PUERTO RICO, INC. | 6945 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Other | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |