BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES

MDR report key: 23735143 · Received December 5, 2025

Report

Report Number
9617032-2025-02151
Event Type
Malfunction
Date Received
December 5, 2025
Date of Event
October 1, 2025
Report Date
February 10, 2026
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED ANY SAMPLES OR PHOTOS FOR INVESTIGATION. A TOTAL OF 100 RETAINED SAMPLES WERE VISUALLY INSPECTED, AND NO ADDITIVE ABNORMALITY OR GEL DEFECTS WERE FOUND. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. COMPLAINTS FOR SAMPLE QUALITY WERE NOT UNDER STATISTICAL CONTROL FOR THE MONTH OF NOVEMBER 2025; THEREFORE, FACTORS THAT MAY CONTRIBUTE TO OIL GEL GLOBULES AND ERRONEOUS RESULTS WERE EVALUATED THROUGH A CLINICAL STUDY TO VERIFY THE DESIGN OF THE DEVICE MET IT¿S INTENDED USE. THERE WERE NO DIFFICULTIES ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES EXHIBITED PROPER FILL. BD WAS UNABLE TO DUPLICATE THE CUSTOMER¿S INDICATED FAILURE MODES (OIL GEL GLOBULES, ERRONEOUS RESULTS-K, NA, CL) VIA CLINICAL INVESTIGATION BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. ALL VISUAL OBSERVATIONS OF BOTH RETENTION AND CONTROL SAMPLES TESTED DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE. REPLICATES OF BOTH RETENTION AND CONTROL SAMPLES WERE ACCEPTABLE IN TERMS OF BOTH PRECISION AND ACCURACY THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODES: OIL GEL GLOBULES, ERRONEOUS RESULTS-K, NA, CL. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. BASED ON AN EVALUATION OF SEVERITY AND FREQUENCY IT WAS DETERMINED THAT NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, GEL PARTICLES AND ERRONEOUS RESULTS WERE SPORADICALLY OBSERVED IN THE SERUM OF AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, GEL PARTICLES AND ERRONEOUS RESULTS WERE SPORADICALLY OBSERVED IN THE SERUM OF AN UNSPECIFIED NUMBER OF DEVICES. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
566614 BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 5076695

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown