FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 23733681 · Received December 5, 2025

Report

Report Number
9617229-2025-21168
Event Type
Injury
Date Received
December 5, 2025
Date of Event
November 17, 2025
Report Date
May 19, 2026
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION WAS RECEIVED ON APRIL 13,2026, WITH LOT NUMBER 3022329. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION (IF APPLICABLE). PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ DEFLATION: OBSERVED AN OPENING THROUGH MICROSCOPIC INSPECTION ASSESSED AS UNIDENTIFIED (TEAR) OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASE AND WEAR ABRASION) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, H6.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". LATER PHYSICIAL REPORTED "WHOLE AT THE SEAL". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232866 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 3022329

Patients

Seq Age Sex Outcome Treatment
1