ULTRASONIC GASTROVIDEOSCOPE
Report
- Report Number
- 3002808148-2025-24258
- Event Type
- Malfunction
- Date Received
- December 4, 2025
- Date of Event
- October 10, 2025
- Report Date
- February 17, 2026
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- ODG
- PMA / PMN Number
- K051541
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE APPROVED FINAL INVESTIGATION. BASED ON THE RESULTS OF THE INVESTIGATION, AND SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION, THE DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED A LITERATURE TITLED ¿ULTRASOUND ENDOSCOPES AND ROUTINE MICROBIOLOGICAL SURVEILLANCE - ANOTHER CRITICAL DEVICE IN REPROCESSING¿. BACKGROUND: PERIODIC ROUTINE MICROBIOLOGICAL SURVEILLANCE OF GASTROINTESTINAL ENDOSCOPES IS RECOMMENDED TO ENSURE ADEQUACY OF REPROCESSING. WE AIMED TO INVESTIGATE WHETHER THERE WERE DIFFERENCES IN CONTAMINATION DEPENDING ON THE TYPE OF ENDOSCOPE. METHODS: WE ANALYZED 437 MICROBIOLOGICAL SURVEILLANCE SAMPLES OBTAINED FROM GASTROINTESTINAL ENDOSCOPES AT A TERTIARY-CARE CENTER IN SWITZERLAND. SAMPLING WAS CARRIED OUT AFTER REPROCESSING IN WASHER-DISINFECTORS ACCORDING TO STANDARD SN EN ISO 15883-4:2018. FOR MICROBIOLOGICAL ANALYSIS, ENDOSCOPE IRRIGATION SOLUTION WAS FILTERED AND TRANSFERRED TO SHEEP BLOOD AGAR. THESE PLATES WERE INCUBATED AT 37°C. IF BACTERIAL GROWTH WAS DETECTED, THE SPECIES WAS IDENTIFIED USING MALDI OR OTHER MICROBIOLOGICAL IDENTIFICATION TECHNIQUES, AND COLONY FORMING UNITS (CFUS) WERE COUNTED. CONTAMINATION WAS DEFINED AS DETECTION OF ANY MICROORGANISM WITH =20 CFUS/20 ML OR DETECTION OF ENTEROBACTERALES, ENTEROCOCCUS SPP., PSEUDOMONAS AERUGINOSA AND OTHER NON-FERMENTERS, STREPTOCOCCUS VIRIDANS AND STAPHYLOCOCCUS AUREUS, INDEPENDENT OF CFU COUNT. RESULTS: TOTALLY, 204 SAMPLES WERE GATHERED FROM 28 GASTROSCOPES, 146 SAMPLES FROM 23 COLONOSCOPES, 45 SAMPLES FROM 5 ULTRASOUND ENDOSCOPES, 27 FROM 4 DUODENOSCOPES AND 15 FROM 3 ENTEROSCOPES, RESPECTIVELY. ELEVEN (2.5%) SAMPLES, GATHERED FROM 5 ENDOSCOPES, INDICATED CONTAMINATION. THE PROPORTION OF SAMPLES INDICATING CONTAMINATION VARIED LARGELY BY ENDOSCOPE TYPE: 20.0% (9/45) OF ULTRASOUND ENDOSCOPES-DERIVED, 1.4% (2/146) OF COLONOSCOPE-DERIVED, 0.5% (1/204) OF GASTROSCOPE-DERIVED, 0% (0/27) OF DUODENOSCOPE-DERIVED AND 0% (0/15) ENTEROSCOPE-DERIVED SAMPLES WERE CONTAMINATED (P<0.001). CONCLUSIONS: OVERALL, A SMALL PROPORTION OF ENDOSCOPES WERE CONTAMINATED. SIGNIFICANT DIFFERENCES WERE OBSERVED BASED ON THE TYPE OF ENDOSCOPE, WITH ULTRASOUND ENDOSCOPES PERFORMING WORST. EVENT 1: ENDOSCOPE SAMPLES INDICATING CONTAMINATION: (1) GIF-1TH190 (EVENT 1) REFER TO TABLE1 AS FOLLOWS (ENDOSCOPE NUMBER 1), DATE (B)(6) 2021, SUCTION CHANNEL, STAPHYLOCOCCUS AUREUS <=1 (CFU/ML), (2) GF-UE160-AL5 (EVENT 1) REFER TO TABLE1 AS FOLLOWS (ENDOSCOPE NUMBER 2), DATE (B)(6) 2021, BIOPSY CHANNEL, CANDIDA PARAPSILOSIS >100000 (CFU/ML), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 25 (CFU/ML), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 55-60 (CFU/ML), DATE (B)(6) 2021, SUCTION CHANNEL, CANDIDA PARAPSILOSIS 25 (CFU/ML), (3) GF-UCT180 (EVENT 1) REFER TO TABLE1 AS FOLLOWS (ENDOSCOPE NUMBER 3), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 20 (CFU/ML), DATE (B)(6) 2021, SUCTION CHANNEL, CANDIDA PARAPSILOSIS 100 (CFU/ML), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 20 (CFU/ML), DATE (B)(6) 2021, SUCTION CHANNEL, CANDIDA PARAPSILOSIS 35 (CFU/ML), (4) GF-UE160-AL5 (EVENT 1) REFER TO TABLE1 AS FOLLOWS (ENDOSCOPE NUMBER 4), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 7000 (CFU/ML), (5) PCF-H190DL (EVENT 1) REFER TO TABLE1 AS FOLLOWS (ENDOSCOPE NUMBER 5), DATE (B)(6) 2021, AIR/WATER CHANNEL, STAPHYLOCOCCUS EPIDERMIS 1 (CFU/ML), ENTEROCOCCUS FAECALIS 1 (CFU/ML). THIS IS REPORT 2 OF 5 FOR: GF-UE160-AL5. DATE (B)(6) 2021, BIOPSY CHANNEL, CANDIDA PARAPSILOSIS >100000 (CFU/ML), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 25 (CFU/ML), DATE (B)(6) 2021, AIR/WATER CHANNEL, CANDIDA PARAPSILOSIS 55-60 (CFU/ML), DATE (B)(6) 2021, SUCTION CHANNEL, CANDIDA PARAPSILOSIS 25 (CFU/ML).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 556104 | ULTRASONIC GASTROVIDEOSCOPE | ULTRASONIC GASTROVIDEOSCOPE | ODG | SHIRAKAWA OLYMPUS CO., LTD. | GF-UE160-AL5 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |