ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2025-00616
- Event Type
- Malfunction
- Date Received
- December 4, 2025
- Date of Event
- September 29, 2025
- Report Date
- January 15, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- UDI-DI
- 00380740186029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. COMPLETED INFORMATION FOR SECTIONS A1 - PATIENT IDENTIFIER/A2 - AGE AT TIME OF EVENT/A2 - AGE UNITS (PATIENT)/A3A - SEX: SAMPLE ID (B)(6)/86/YEARS/FEMALE, SAMPLE ID (B)(6)/23/YEARS/FEMALE, SAMPLE ID (B)(6)/50/YEARS/FEMALE, SAMPLE ID (B)(6)/37/YEARS/FEMALE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P13, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 04Z21, WITH 510K/PMA/BLA NUMBER K202525.
DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A REVIEW OF TRACKING AND TRENDING FOR THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT DID NOT IDENTIFY ANY TRENDS RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, NO TRENDS WERE IDENTIFIED FOR LOT 77409UD00. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT LOT NUMBER OR ISSUE. THE OVERALL PERFORMANCE OF THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENTS IN THE FIELD WAS REVIEWED USING DATA GATHERED FROM CUSTOMERS WORLDWIDE. THE REVIEW OF FIELD DATA FOR LOTS 77408UD00 AND 77406UD00, WHICH CONTAINS THE SAME BULK MATERIAL AS COMPLAINT LOT 77409UD00 DETERMINED THE MEDIAN PATIENT RESULTS ARE WITHIN THE ESTABLISHED LIMITS AND COMPARABLE TO THE HISTORICAL REAGENT LOT PERFORMANCE. LABELING WAS REVIEWED AND WAS FOUND TO ADEQUATELY ADDRESS THE CURRENT ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED FOR THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT, LOT NUMBER 77409UD00.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULTS GENERATED FOR FEMALE PATIENT SAMPLES. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE - FEMALE <16 NG/L; MALE <26 NG/L): SAMPLE ID (B)(6), (86-YEAR-OLD) (B)(6) 2025, INITIAL RESULT ON (B)(6) = 60 NG/L, REPEAT RESULT = 2 NG/L (<3 NG/L), NEW SAMPLE OBTAINED AND RESULT = <3 NG/L. SAMPLE ID (B)(6), (23-YEAR-OLD) (B)(6) 2025, INITIAL RESULT ON (B)(6) = 93 NG/L, (B)(6) 2025 SAMPLE ID (B)(6) RESULT ON (B)(6) = 2 NG/L (<3 NG/L). SAMPLE ID (B)(6), (50-YEAR-OLD) (B)(6) 2025 INITIAL RESULT ON (B)(6) = 41 NG/L, REPEAT RESULT = (<3 NG/L), NEW SAMPLE OBTAINED SAMPLE ID (B)(6), RESULT ON (B)(6) = 0 NG/L (<3 NG/L), 3RD SAMPLE OBTAINED AND RESULT = <3 NG/L. SAMPLE ID (B)(6), (37-YEAR-OLD) (B)(6) 2025, INITIAL RESULT ON (B)(6) = 121 NG/L, REPEAT RESULT = <3 NG/L, (B)(6) 2025 NEW SAMPLE OBTAINED SAMPLE ID(B)(6), RESULT ON (B)(6) = 0 NG/L (<3 NG/L) . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULTS GENERATED FOR FEMALE PATIENT SAMPLES. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE - FEMALE <16 NG/L; MALE <26 NG/L): SAMPLE ID (B)(6) (86-YEAR-OLD) (B)(6) 2025 INITIAL RESULT ON (B)(6) = 60 NG/L, REPEAT RESULT = 2 NG/L (<3 NG/L), NEW SAMPLE OBTAINED AND RESULT = <3 NG/L SAMPLE ID (B)(6) (23-YEAR-OLD) (B)(6) 2025 INITIAL RESULT ON (B)(6) = 93 NG/L, (B)(6) 2025 SAMPLE ID (B)(6) RESULT ON (B)(6) = 2 NG/L (<3 NG/L). SAMPLE ID (B)(6) (50-YEAR-OLD) (B)(6) 2025 INITIAL RESULT ON (B)(6) = 41 NG/L, REPEAT RESULT = (<3 NG/L), NEW SAMPLE OBTAINED SAMPLE ID (B)(6) RESULT ON AI30171 = 0 NG/L (<3 NG/L), 3RD SAMPLE OBTAINED AND RESULT = <3 NG/L. SAMPLE ID M(B)(6) (37-YEAR-OLD) (B)(6) 2025 INITIAL RESULT ON (B)(6) = 121 NG/L, REPEAT RESULT = <3 NG/L, (B)(6) 2025 NEW SAMPLE OBTAINED SAMPLE ID (B)(6) RESULT ON (B)(6) = 0 NG/L (<3 NG/L). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2807800 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 77409UD00 | 00380740186029 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6) |