FDA Adverse Event Malfunction Summary report: N

FISHER & PAYKEL HEALTHCARE

MDR report key: 23711686 · Received December 3, 2025

Report

Report Number
9611451-2025-01755
Event Type
Malfunction
Date Received
December 3, 2025
Report Date
January 9, 2026
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). SECTION G4: THE IW990AEA WALL MOUNT INFANT WARMER IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K971441. METHOD: THE COMPLAINT DEVICE IW990AEA WALL MOUNT INFANT WARMER WAS NOT RECEIVED AT FISHER & PAYKEL HEALTHCARE (F&P). OUR INVESTIGATION IS THUS BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY AND OUR KNOWLEDGE OF THE PRODUCT. RESULT: THE HEALTHCARE FACILITY REPORTED THAT THE IW990AEA WALL MOUNT INFANT WARMER WAS SHOWING ERROR CODE 17. CONCLUSION: THE REPORTED EVENT COULD HAVE CAUSED DUE TO EITHER HARNESS DEGRADATION OR THE ELEMENT BEING A OPEN CIRCUIT. THE USER MANUAL FOR THE IW990 INFANT WARMER STATES THE FOLLOWING: "ENSURE ALL WARMER PARTS AND ACCESSORIES ARE CHECKED BEFORE RETURNING THE DEVICE TO SERVICE." "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYDROCHLORITES, PEROXIDES, GLUTARALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE." "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES." THE DEVICE TECHNICAL/SERVICE MANUAL ALSO CONTAINS A CHECKLIST WHICH SPECIFIES THAT USERS PERFORM SAFETY, PERFORMANCE, AND FUNCTIONAL CHECKS AT LEAST ONCE A YEAR.

Additional Manufacturer Narrative · 0

(B)(6). SECTION G4: THE IW990AEA WALL MOUNT INFANT WARMER IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K971441. FISHER & PAYKEL HEALTHCARE (F&P) IS IN THE PROCESS OF COMPLETING OUR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN AUSTRALIA REPORTED THAT THE IW990AEA INFANT WARMER WALLMOUNT NONSRVO AEA IS SHOWING ERROR CODE 17. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN AUSTRALIA REPORTED THAT THE IW990AEA INFANT WARMER WALLMOUNT NONSRVO AEA IS SHOWING ERROR CODE 17. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1979580 FISHER & PAYKEL HEALTHCARE INFANT WARMER FMT FISHER & PAYKEL HEALTHCARE LTD IW990AEA 070620

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown