FDA Adverse Event Injury Summary report: N

VOALTE NURSE CALL

MDR report key: 23707309 · Received December 3, 2025

Report

Report Number
2027454-2025-00236
Event Type
Injury
Date Received
December 3, 2025
Date of Event
November 4, 2025
Report Date
December 3, 2025
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
ILQ
UDI-DI
00887761995079
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H11: A REVIEW OF THE DEVICE LOG INDICATED THAT NO STAFF EMERGENCY CALLS WERE PLACED DURING THAT TIME. UPON INSPECTION OF THE DEVICE, IT WAS FOUND THE EMERGENCY SWITCH WAS DIRTY INSIDE CAUSING THE CANCEL BUTTON TO BE STUCK IN; HOWEVER, THE DEVICE IS BEING RETURNED TO BAXTER AND THE INVESTIGATION IS STILL ONGOING. THE INVESTIGATION IS CURRENTLY ONGOING, A FINAL REPORT WILL BE SUBMITTED UPON COMPLETION.

Description of Event or Problem · 0

IT WAS REPORTED A HOSPITAL PAGING SYSTEM FAILED AND A PATIENT PASSED AWAY. IN THE EMERGENCY ROOM AN EMERGENCY CALL FAILED WITH A PATIENT ALLEGEDLY REQUIRING ¿EXTRICATION FROM THE CAR AND BEGINNING ADVANCED LIFE SUPPORT WAS DELAYED¿. THE CUSTOMER STATED, ¿THE PATIENT IS DECEASED, AND THEY CANNOT STATE IF IT WAS DUE TO THE DELAY¿. THE AMOUNT OF TIME THAT THE CARE WAS ¿DELAYED¿ WAS NOT PROVIDED. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL DETAILS HAVE BEEN UNSUCCESSFUL. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2829041 VOALTE NURSE CALL SYSTEM, COMMUNICATION, POWERED ILQ BAXTER HEALTHCARE CORPORATION P2599N NA 00887761995079

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Death NI.