FDA Adverse Event Malfunction Summary report: N

LEAD

MDR report key: 23707029 · Received December 3, 2025

Report

Report Number
MW5179837
Event Type
Malfunction
Date Received
December 3, 2025
Report Date
October 1, 2025
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
NVY
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MN, US
Reporter Occupation
OTHER
Health Professional
*

Narratives

Description of Event or Problem · 0

Q: IS THERE A CONCERN WITH THIS LEAD WITH LOW PACING IMPEDANCE VALUES AND LOW R WAVES SINCE IT IS ON ADVISORY WITH BOSTON SCIENTIFIC? D: BOSTON SCIENTIFIC 0292 ENDOTAK SINGLE COIL LEAD WITH IMPEDANCES AROUND 68 OHMS IMPLANTED IN 2017. IMPEDANCES AT TIME OF IMPLANT ARE NOT KNOWN. R: NOTED TO CALLER THAT THE BOSTON SCIENTIFIC ADVISORY IS RELATED TO HIGH/RISING DEFIB IMPEDANCE VALUES. THIS LEAD SHOWS LOW TTC PACING IMPEDANCES (190 OHMS IMPEDANCES SINCE AT LEAST 2022). IMPEDANCE VALUES AND R WAVES HAVE REMAINED STABLE BUT LOW IN THIS INTEGRATED BIPOLAR LEAD WITH NO RING ELECTRODE. DEFIB IMPEDANCE SHOWS 66 OHMS. THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2829323 LEAD PERMANENT DEFIBRILLATOR ELECTRODES NVY BOSTON SCIENTIFIC CORPORATION 0292

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown