FDA Adverse Event Malfunction Summary report: N

VP 2000 PROCESSOR

MDR report key: 2370401 · Received April 4, 2011

Report

Report Number
3005248192-2011-00002
Event Type
Malfunction
Date Received
April 4, 2011
Date of Event
March 7, 2011
Report Date
April 4, 2011
Manufacturer
ABBOTT MOLECULAR, INC.
Product Code
KEY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ABBOTT MOLECULAR REQUESTED THE RETURN OF THE INSTRUMENT WITH THE ALLEGED SHARP EDGES FOR COMPLAINT INVESTIGATION. THE COMPLAINT IS CURRENTLY UNDER INVESTIGATION. ABBOTT MOLECULAR COMPLETED A SHARPNESS TEST ON A VP2000 INSTRUMENT CURRENTLY IN USE AT ABBOT MOLECULAR. THIS INSTRUMENT PASSED ON ALL USER ACCESSIBLE EDGES (PER UL STANDARD 1439).

Description of Event or Problem · 1

THE ABBOTT VP 2000 PROCESSOR IS A DEVICE DESIGNED TO AUTOMATE AND STANDARDIZE SLIDE SPECIMEN PROCESSING AND ROUTINE SLIDE STAINING FOR THE LABORATORY. A CUSTOMER AND A FIELD SERVICE REPRESENTATIVE IN (B)(4) ALLEGED THAT THE METAL EDGES OF AN ABBOTT VP 2000 PROCESSOR (HEATER BLOCK, COVERS) UNIT ARE VERY SHARP. THE PARTS WITH EDGES THAT ARE ALLEGEDLY SHARP INCLUDE THE COVER FOR THE AMBIENT TEMPERATURE BATHS AND WHEN THE HEATED BASINS ARE REMOVED, THE OPENINGS OF THE HEATER REAGENT BASINS. DRYING STATION AND WATER BATH HAVE ALLEGEDLY SHARP EDGES ALSO. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VP 2000 PROCESSOR STAINER, TISSUE, AUTOMATED KEY ABBOTT MOLECULAR, INC.

Patients

Seq Age Sex Outcome Treatment
1