FDA Adverse Event Malfunction Summary report: N

CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM

MDR report key: 23702791 · Received December 3, 2025

Report

Report Number
3003120897-2025-00669
Event Type
Malfunction
Date Received
December 3, 2025
Date of Event
November 10, 2025
Report Date
February 6, 2026
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC.
Product Code
HWE
UDI-DI
00643169754584
PMA / PMN Number
K170679
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS #(B)(6) :PART # 6550017 : LOT # K23K1190 VISUAL INSPECTION CONFIRMED THE END OF THE TAB EXTENDER THAT INTERFACES WITH THE BREAK OFF SCREW HAS BROKEN. THE DAMAGE TO THE EXTENDER IS CONSISTENT WITH OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL REGARDING AN EXTENDER. IT WAS REPORTED THAT THE EXTENDER WOULD NOT CONNECT WITH THE SCREW. THERE WAS NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2848744 CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED M HWE MEDTRONIC SOFAMOR DANEK USA, INC. 6550017 K23K1190 00643169754584

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown