FDA Adverse Event Malfunction Summary report: N

ECHO TABLE

MDR report key: 2369811 · Received December 8, 2011

Report

Report Number
1932056-2011-00007
Event Type
Malfunction
Date Received
December 8, 2011
Date of Event
September 23, 2011
Report Date
October 19, 2011
Manufacturer
MEDICAL POSITIONING, INC.
Product Code
DPS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IT WAS DISCOVERED THROUGH OUR INVESTIGATION THAT THE LATCHES WHICH HOLD THE DROP-SECTION CLOSED HAD COME OUT OF ADJUSTMENT, THE PROBLEM WITH THE CUSTOMER'S DROP-SECTION LATCHES WAS PROMPTLY REMEDIED BY ADJUSTING THE LATCHES. WITH AN INSTALLED PRODUCT BASE OF OVER 5000 UNITS, A SIMILAR SITUATION HAS NEVER BEEN REPORTED TO THE CO. THE USER MANUAL SECTION FOR PREVENTIVE MAINTENANCE INSTRUCTS THE CUSTOMER TO INSPECT THE DROP SECTION LATCH MECHANISM ANNUALLY AND ADJUST IF NECESSARY. THE COMPANY WILL CONTINUE TO MONITOR CUSTOMER COMPLAINTS FOR OUT OF ADJUSTMENT DROP-SECTION LATCHES TO SEE IF ANY CORRECTIVE ACTION IS NECESSARY.

Description of Event or Problem · 1

THE DROP-SECTION ON AN ECHO TABLE MODEL 2272 WILL FALL OPEN WHEN PT WEIGHT IS APPLIED. NO INJURY OR DEATH WAS REPORTED AS A RESULT OF THE COMPLAINT FILED WITH MEDICAL POSITIONING, INC. BUT IT IS FORESEEABLE THAT IF PT WEIGHT IS APPLIED TO THE DROP-SECTION DURING PT INGRESS/EGRESS, IT MAY LEAD TO PT FALL WHICH COULD CAUSE A SERIOUS INJURY SUCH AS A BREAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHO TABLE ECHOCARDIOGRAPHY PROCEDURE TABLE DPS MEDICAL POSITIONING, INC. 2272

Patients

Seq Age Sex Outcome Treatment
1 Other