FDA Adverse Event Death Summary report: N

DOUBLE TRACHEOSTOMY SWIVEL ADAPTER WITH SUCTION PORT

MDR report key: 23692 · Received July 25, 1995

Report

Report Number
MW1006563
Event Type
Death
Date Received
July 25, 1995
Date of Event
May 29, 1995
Report Date
July 21, 1995
Manufacturer
MEDICOMP
Product Code
BZA
Adverse Event
Yes
Report Source
Voluntary report
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT WAS CONNECTED TO VENTILATOR. NURSE WENT IN TO CHECK ON PT. PT HAD EXPIRED. IT WAS NOTICED SWIVEL WAS DISCONNECTED FROM TRACHEOSTOMY TUBE. UNIT DID NOT ALARM. SWIVEL PIECE CONNECTS VENTILATOR HOSE TO TRACHEOSTOMY PIECE. RISK MANAGER STATED THEY HAVE NOTICED A TREND WITH SWIVEL PIECE. WHEN VENTILATOR IS IN MODE S INV, IT DOES NOT ALARM. WHEN VENTILATOR IS IN AC MODE, IT DOESN'T ALARM. THE PROBLEM LIES WITH THE DESIGN OF THE SWIVEL. IT SEEMS THAT BACK PRESSURE IS PRODUCED AND IT CAUSES THE PIECE TO DISCONNECT LEAVING VENTILATED PT WITHOUT AN OXYGEN SUPPLY.FOLLOW-UP INFORMATION RECEIVED 8/15/95: TRACHEAL SWIVEL DUE TO DESIGN FAULT DOES NOT ALLOW THE VENTILATOR TO ALARM WHEN THE TUBING GETS DISCONNECTED FROM THE TRACHEAL TUBE. THE SWIVEL RETAINS ENOUGH PRESSURE OR BACK PRESSURE THAT THE VENTILATOR SENSES THE PRESSURE AND DOES NOT ALARM. THIS OCCURS WHILE THE VENTILATOR IS IN THE AC MODE OF OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DOUBLE TRACHEOSTOMY SWIVEL ADAPTER WITH SUCTION PORT SWIVEL ADAPTER BZA MEDICOMP MC 4475 6350

Patients

Seq Age Sex Outcome Treatment
1 42 YR Death