IMP TM 3.7MM MTX,11.5MM
Report
- Report Number
- 0001038806-2025-03451
- Event Type
- Injury
- Date Received
- December 1, 2025
- Date of Event
- October 27, 2023
- Report Date
- January 22, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- UDI-DI
- 00889024018877
- PMA / PMN Number
- K132258
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4).
ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: A. PATIENT IDENTIFIER A. AGE AT TIME OF EVENT A. WEIGHT UPDATED TO UNKNOWN A. GENDER B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE ZIMVIE RECEIVED ONE (1) TMMB11, (IMP TM 3.7MM MTX,11.5MM) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE. THE IMPLANT WAS OBSERVED FRACTURED IN HALF, AND THE REMAINING FRACTURED PART WAS NOT RETURNED FOR EVALUATION. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1230588. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1230588 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿FRACTURE IMPLANT¿ BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE PARAFUNCTIONAL HABITS/PATIENT FACTORS, OR CUSTOMER ERROR. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
THE DOCTOR REPORTS THAT THE DENTAL IMPLANT LOCATED IN POSITION NUMBER 13 FAILED BECAUSE IT FRACTURED AT HALF. IMPLANT WAS REMOVED. THE DOCTOR REPORTS THAT THE PROCEDURE WAS NOT CONCLUDED BY PLACING ANOTHER IMPLANT.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12841 | IMP TM 3.7MM MTX,11.5MM | DENTAL IMPLANT | DZE | ZIMVIE US CORP LLC | 1230588 | 00889024018877 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Male | Required Intervention |