HEART LUNG MACHINE
Report
- Report Number
- 8010762-2025-0000528
- Event Type
- Malfunction
- Date Received
- December 1, 2025
- Date of Event
- November 12, 2025
- Report Date
- December 16, 2025
- Manufacturer
- MAQUET CARDIOPULMONARY GMBH
- Product Code
- KFM
- UDI-DI
- 04037691718910
- PMA / PMN Number
- K991864
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INVESTIGATION IS ONGOING. FURTHER INFORMATION HAS BEEN REQUESTED BUT HAS NOT YET BEEN RECEIVED. A FOLLOW UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. NOTE: THIS EVENT OCCURRED ON THE CHINESE MARKET. IT IS A SIGNIFICANTLY SIMILAR DEVICE TO "BASE UNIT, ROTAFLOW CONSOLE¿ WHICH IS SOLD IN THE USA UNDER PREMARKET SUBMISSION NUMBER: K991864. FOR SECTION D4 ALL AVAILABLE IDENTIFYING INFORMATION IS FOR THE OUT OF THE US DEVICE, SUBJECTED TO THIS REPORT. THEREFORE, NO GUDID INFORMATION EXISTS. FURTHERMORE, UDI INFORMATION (PRIMARY DI NUMBER) PROVIDED IN D4 IS FOR ROTAFLOW CONSOLE WITH CATALOG NUMBER: 701046405.
THE EVENT OCCURRED IN CHINA. IT WAS REPORTED THAT THE ERROR MESSAGE ¿HEAD ERROR¿ OCCURRED ON THE ROTAFLOW CONSOLE. THE FAILURE OCCURRED DURING USE AND THE DEVICE WAS REPLACED IMMEDIATELY UPON OCCURRENCE. NO HARM TO ANY PERSON HAS BEEN REPORTED. THE REPORTED ¿HEAD ERROR¿ COULD LEAD TO A PUMP STOP DURING USE, THEREFORE A REPORT IS REQUIRED. THE PATIENT DATA WAS NOT SHARED BY CUSTOMER. A GETINGE FIELD SERVICE TECHNICIAN INVESTIGATED THE AFFECTED ROTAFLOW CONSOLE (RFC) WITH S/N (B)(6) AND THE ROTAFLOW DRIVE (RFD) WITH S/N (B)(6). THE REPORTED "HEAD ERROR" COULD NOT BE REPRODUCED DURING STARTUP TESTING, AND SIMULATED OPERATION PROCEEDED NORMALLY. IT IS RECOMMENDED TO PERFORM A PREVENTATIVE REPLACEMENT OF THE CORRESPONDING COMPONENTS BEFORE RESUMING USE. NO PARTS WERE REPLACED YET, AS THE CUSTOMER HAS NOT ACCEPTED THE PREVENTATIVE PART REPLACEMENT. THE DEVICE IS CURRENTLY OUT OF SERVICE. BASED ON THESE INVESTIGATION RESULTS THE REPORTED FAILURE "HEAD ERROR" COULD NOT BE CONFIRMED. HOWEVER, THE MOST PROBABLE ROOT CAUSE FOR THE REPORTED FAILURE "HEAD ERROR" COULD BE DETERMINED AS: 1. THE HEAD ERROR FOLLOWS THE SIG ERROR. WHEN THE ULTRASONIC CREAM IS APPLIED (DUE TO THE SIG ERROR) TO THE FLOW BUBBLE SENSOR AND THE DISPOSABLE IS MOVED CLOSE TO THE ROTA FLOW DRIVE, THE MAGNETS IN THE CENTRIFUGAL PUMP INTERFERE WITH THE SENSORS IN THE ROTA FLOW DRIVE. THIS ERROR CAN BE OVERWRITTEN BY RESTARTING THE ROTA FLOW CONSOLE. 2. IF THE RPM / LPM IS SET TO ZERO AND THE DRIVE AS WELL AS THE INSERTED CENTRIFUGAL PUMP IS SHAKEN THIS CAUSES MAGNETIC UNCOUPLING OF THE CENTRIFUGAL PUMP AND THUS LEADS TO THE HEAD ERROR. THIS ERROR CAN BE OVERWRITTEN BY RESTARTING THE ROTA FLOW CONSOLE. ROTAFLOW CONSOLE: THE REVIEW OF THE NON-CONFORMITIES HAS BEEN PERFORMED ON (B)(6) 2025 FOR THE PERIOD OF 2020-10-28 TO 2025-11-18. IT DOES NOT SHOW ANY NON-CONFORMITY IN REGARD TO THE REPORTED PRODUCT AND FAILURE. THERE IS NO INDICATION THAT MANUFACTURING ISSUES OCCURRED DURING THIS TIME, THUS PRODUCTION RELATED INFLUENCES ARE UNLIKELY. THE ROTAFLOW CONSOLE WITH S/N (B)(6) WAS PRODUCED IN (B)(6) 2020. ROTAFLOW DRIVE: THE REVIEW OF THE NON-CONFORMITIES HAS BEEN PERFORMED ON (B)(6) 2025 FOR THE PERIOD OF 2020-10-14 TO 2025-11-18. IT DOES NOT SHOW ANY NON-CONFORMITY IN REGARD TO THE REPORTED PRODUCT AND FAILURE. THERE IS NO INDICATION THAT MANUFACTURING ISSUES OCCURRED DURING THIS TIME, THUS PRODUCTION RELATED INFLUENCES ARE UNLIKELY. THE ROTAFLOW DRIVE WITH S/N (B)(6) WAS PRODUCED IN (B)(6) 2020. IN ADDITION, A REVIEW FOR POTENTIAL FIELD ACTIONS AND CAPAS RELATED TO THE FAILURE AND PRODUCT IN THIS COMPLAINT WAS PERFORMED. THIS COMPLAINT IS NOT IN SCOPE OF ANY ONGOING FIELD ACTIONS AND/OR CAPAS. FOR THE AFFECTED SERIAL NUMBERS WITHIN THE TIMEFRAME OF THE SEARCH NO ADDITIONAL COMPLAINT WAS FOUND FOR THE SAME ISSUE. IN ORDER TO AVOID REOCCURRENCE OF THE REPORTED FAILURE "HEAD ERROR", IT IS NECESSARY TO FOLLOW THE CHAPTER IN THE INSTRUCTION FOR USE HEART-LUNG SUPPORT SYSTEM ROTAFLOW SYSTEM/ 4.4 / EN / 15. CHAPTER 4.1.3: SWITCH OFF THE ROTAFLOW CONSOLE ON/OFF SWITCH BEFORE CONNECTING THE ROTAFLOW DRIVE TO OR DISCONNECTING IT FROM THE ROTAFLOW CONSOLE. OTHERWISE, THE ROTAFLOW CONSOLE MAY BE DAMAGED. CHAPTER 2.2.3. TAKE DAMAGED DEVICES OUT OF SERVICE IMMEDIATELY AND HAVE THEM TESTED BY THE AUTHORIZED SERVICE PERSONNEL. IT IS ALSO NECESSARY TO FOLLOW THE CHAPTER IN THE SERVICE MANUAL FOR USE HEART-LUNG SUPPORT SYSTEM ROTAFLOW SYSTEM/ EN / 03 CHAPTER 1.7 SERVICE-RELATED WORK ON A DEVICE MAY ONLY BE CARRIED OUT BY SERVICE TECHNICIANS WHO HAVE BEEN TRAINED AND INSTRUCTED AND CERTIFIED BY GETINGE. SERVICE-RELATED WORK ON A DEVICE CARRIED OUT BY UNQUALIFIED SERVICE TECHNICIANS' MAY LED TO INJURY OF THE SERVICE TECHNICIAN, PATIENT OR OTHER PERSONS OR MAY LED TO DEVICE DAMAGE. CHAPTER 1.10 A REPAIR IS CARRIED OUT FOR MAINTENANCE AFTER DAMAGE OR MALFUNCTION OF A ROTAFLOW SYSTEM. THE OCCURRENCE RATE WAS CALCULATED FOR THE REPORTED ISSUE AND IT WAS DETERMINED THAT THIS IS NOT A SYSTEMIC ISSUE. THEREFORE, NO REMEDIAL ACTION IS REQUIRED. THE OCCURRENCE RATE RELATED TO THE REPORTED ISSUE IS CURRENTLY BEING MONITORED AS PART OF MAQUET CARDIOPULMONARY¿ S TRENDING PROGRAM AND ADDITIONAL INVESTIGATIONS OR CORRECTIONS WILL BE IMPLEMENTED IN CASE OF ADVERSE TRENDING.
THE EVENT OCCURRED IN CHINA. IT WAS REPORTED THAT THE ERROR MESSAGE ¿HEAD ERROR¿ OCCURRED ON THE ROTAFLOW CONSOLE. THE FAILURE OCCURRED DURING USE AND THE DEVICE WAS REPLACED IMMEDIATELY UPON OCCURRENCE. NO HARM TO ANY PERSON HAS BEEN REPORTED. THE REPORTED ¿HEAD ERROR¿ COULD LEAD TO A PUMP STOP DURING USE, THEREFORE A REPORT IS REQUIRED. COMPLAINT ID: (B)(4).
COMPLAINT ID: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2727772 | HEART LUNG MACHINE | PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE | KFM | MAQUET CARDIOPULMONARY GMBH | MCP00706291#ROTAFLOW PUMP MODULE | 04037691718910 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |