INVESTIGATION RESULTS: ONE PHOTO WAS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. IN THE PHOTO, A SYRINGE CONTAINS A WHITE LIQUID WITH A BLACK-COLORED PARTICLE; THEREFORE, FOREIGN MATTER IN THE SYRINGE CAN BE VERIFIED. ALTHOUGH THE PHOTO SHOWS WHAT COULD BE A PARTICLE OR FOREIGN MATTER, THE ABSENCE OF THE PHYSICAL SAMPLE MAKES IT IMPOSSIBLE PERFORM IDENTIFICATION ANALYSES TO DETERMINE THE ORIGIN OR NATURE OF THE MATERIAL. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH. BASED ON THE QUALITY TEAM'S INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED WITHOUT A PHYSICAL SAMPLE. THIS INCIDENT HAS BEEN ADDED TO OUR DATABASE OF REPORTED INCIDENTS.
B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. E.1. ADDRESS WAS NOT LOCATED AND IL WAS USED. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
NO ADDITIONAL INFORMATION.
IT WAS REPORTED THAT THE BD SYRINGE 5ML LL BNS HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL#: 304656 BATCH#: 4205617 IT WAS REPORTED BY THE CUSTOMER THAT BLACK FOREIGN OBJECT IN SYRINGE. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. MEDLINE COMPLAINT #:(B)(4). DEFECT DESCRIPTION: CUSTOMER REPORT: ¿BLACK FOREIGN OBJECT IN SYRINGE.¿ XXXX KIT #: DYNJRA1665B XXXXXX PART #: 153245 PRODUCT DESCRIPTION: SYRINGE 5ML LL BNS VENDOR PART #: 304656 LOT #: 4205617 DATE REPORTED: 10/28/2025 SAMPLE RECEIVED: NO RESPONSE NEEDED: COMPLETION OF THE ATTACHED SOP31 FORM WITHIN 15 DAYS OF RECEIPT 304656 4205617 ADDITIONAL INFORMATION PROVIDED: "I WANT TO CLARIFY THAT THE PARTICULATE ISN¿T COMING FROM THE MEDICATION¿THEY NO LONGER ENCOUNTERED ISSUES WHEN THEY USED A DIFFERENT SYRINGE THAT WASN¿T SUPPLIED IN THE KIT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51834 | BD SYRINGE 5ML LL BNS | PISTON SYRINGE | FMF | BECTON DICKINSON DE MEXICO | 4205617 | 00382903046560 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |