ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-09186
- Event Type
- Malfunction
- Date Received
- December 12, 2011
- Report Date
- November 16, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT BUT THE KEYPAD SYMBOLS WERE WORN; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A SCRATCHED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE REPORTER STATED THAT ALL OF THE KEYPAD BUTTONS HAD A DELAYED RESPONSE FOR THE PAST SIX WEEKS AND REQUIRED SEVERAL PRESSES ON THE KEYPAD IN ORDER TO ELICIT A RESPONSE. THE REPORTER ALSO STATED THAT THE DOWN ARROW BUTTON HAD NORMAL SPRING BACK BUT REQUIRED HARD PRESSES IN ORDER FOR IT TO RESPOND. THE PUMP WAS REPORTEDLY CLEANED WITH WATER USING A SOFT DAMP CLOTH OR Q-TIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |