PICCOLO XPRESS CHEMISTRY ANALYZER
Report
- Report Number
- 2939693-2025-00005
- Event Type
- Malfunction
- Date Received
- November 26, 2025
- Date of Event
- October 2, 2025
- Report Date
- November 26, 2025
- Manufacturer
- ABAXIS, INC.
- Product Code
- JJG
- UDI-DI
- EABA110000001
- PMA / PMN Number
- K934592
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- 003
Narratives
02OCT2025/ UNKNOWN TIME: A WHOLE BLOOD (WB) SAMPLE WAS ANALYZED USING THE PICCOLO XPRESS ANALYZER SERIAL NUMBER: (B)(6), AND THE PICCOLO CMP PANEL, LOT: 5222BC1. RESULTS: CREATININE (CRE)= 4.3* MG/DL (REFERENCE RANGE: 0.6-1.2 MG/DL). RERUN: 02OCT2025/UNKNOWN TIME: THE SAME SAMPLE FROM THE PREVIOUS RUN WAS SENT TO AN EXTERNAL LABORATORY, AND PLASMA WAS ANALYZED USING AN ABBOTT ALINITY. RESULTS: CREATININE (CRE)= 0.73 MG/DL (REFERENCE RANGE: 0.7-1.5MG/DL). REPEAT: 02OCT2025/ UNKNOWN TIME: A WHOLE BLOOD (WB) SAMPLE WAS ANALYZED USING THE PICCOLO XPRESS ANALYZER SERIAL NUMBER: (B)(6) PICCOLO CMP PANEL, LOT: 5222BC1. RESULTS: CREATININE (CRE)=0.9 MG/DL (REFERENCE RANGE: 0.6-1.2 MG/DL). ZOETIS REQUESTED FURTHER CLARIFICATION OF THE COMPLAINT DETAILS. THE PATIENT RECEIVED TREATMENT; HOWEVER, THE SPECIFIC TREATMENT ADMINISTERED, AND THE TIME OF ADMINISTRATION ARE UNKNOWN. A RETURN MATERIAL AUTHORIZATION (RMA) HAS BEEN RECOMMENDED FOR FURTHER INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.
29OCT2025: ZOETIS UNION CITY TECHNICAL SUPPORT RECEIVED A CALL FROM A HEALTHCARE PROFESSIONAL REPORTING AN UNEXPECTED HIGH CREATININE (CRE) RESULT USING THE PICCOLO XPRESS ANALYZER, SERIAL NUMBER: (B)(6), AND THE PICCOLO COMPREHENSIVE METABOLIC PANEL (CMP) LOT: 5222BC1. THE PATIENT WAS TREATED. CHEM HIGH/LOW: HIGH, ERROR CODES: CHEM CRE / CREAT PROBLEM OR QUESTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1912907 | PICCOLO XPRESS CHEMISTRY ANALYZER | PICCOLO XPRESS CHEMISTRY ANALYZER | JJG | ABAXIS, INC. | EABA110000001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Female |