FDA Adverse Event Injury Summary report: N

DAVINCI XI

MDR report key: 23662352 · Received November 26, 2025

Report

Report Number
2955842-2025-46113
Event Type
Injury
Date Received
November 26, 2025
Date of Event
October 30, 2025
Report Date
November 26, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE VISITED THE SITE AND REPLACED THE IESU GENERATOR DUE TO ERROR. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE IESU GENERATOR WAS ANALYZED AND FOUND THE REPORTED ISSUE WAS CONFIRMED WITH MULTIPLE U-04 ERRORS, ALONG WITH M-11 AND 2-8A ERRORS. INSPECTION FOUND RELATED ISSUES BUT COULD NOT REPLICATE THE EVENT. IESU LOGS SHOWED A C-00 ERROR CONSISTENT WITH SYSTEM LOGS. THE UNIT PASSED ALL RECOGNITION AND ENERGY TESTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED BENIGN HYSTERECTOMY SURGICAL PROCEDURE, THE CUSTOMER REPORTED TO INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) THAT THEY WERE NOT ABLE TO FIRE ENERGY, AND THE CORDS WERE NOT WORKING. THE TSE CONFIRMED ERROR U-04 ERROR IN LOGS AND HAD CUSTOMER POWER CYCLE INTEGRATED ELECTROSURGICAL UNIT (IESU). THE CUSTOMER DOCKED AND UNDOCKED TO FIRE ENERGY; HOWEVER, ISSUE PERSISTED. THE TSE RECOMMENDED TO TRY ANOTHER INSTRUMENT; HOWEVER, THE CUSTOMER WAS UNABLE TO FURTHER TROUBLESHOOT. THE PROCEDURE WAS ABORTED POST-ANESTHESIA AND PORTS PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1912891 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-43 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention DA VINCI INSTRUMENTS AND ACCESSORIES.