LIPOASPIRATE WASH SYSTEM
Report
- Report Number
- 1651189-2025-09848
- Event Type
- Injury
- Date Received
- November 26, 2025
- Date of Event
- November 5, 2025
- Report Date
- November 10, 2025
- Manufacturer
- TIGER AESTHETICS MEDICAL
- Product Code
- QKL
- PMA / PMN Number
- BK190433
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- 003
Narratives
TIGER AESTHETICS MEDICAL COMPLAINT NUMBER: (B)(4). DEVICE EVALUATION: G3, G6, H2, H3, H6, H10. TIGER AESTHETICS MEDICAL PERFORM AN INVESTIGATION AND FOUND THE FAILURE TO GENERATE SUCTION WAS CAUSED BY AN IMPROPERLY SEATED DRAWER GASKET SEAL DURING OR SETUP. NO MANUFACTURING DEFECT WAS IDENTIFIED, AND NO TREND OR SYSTEMIC ISSUE EXISTS. THIS COMPLAINT IS CONSIDERED USE-RELATED EVENT WITH LOW RISK AND NO PATIENT HARM. NO FURTHER ACTION REQUIRED BEYOND STANDARD MONITORING. TIGER AESTHETICS MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
TIGER AESTHETICS MEDICAL COMPLAINT #: (B)(4). PATIENT AGE DEFAULTED TO "99" AS NOT PATIENT INFORMATION WAS PROVIDED. AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. TIGER AESTHETICS MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER REPORTED TO TIGER AESTHETICS MEDICAL THAT THEY HAD TROUBLE GETTING SUCTION FROM THE UNIT AT THE START OF LIPOSUCTION. IT WAS FOUND THAT THE TRAY THAT IS PULLED OUT LAST IN THE PROCESSING PROCEDURE WAS NOT SEATED CORRECTLY. AN ADDITIONAL VIALITY UNIT WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2137954 | LIPOASPIRATE WASH SYSTEM | FAT TRANSFER | QKL | TIGER AESTHETICS MEDICAL | VIALITY-1400 | A57389 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 99 YR | Unknown |